Preparation
§ Select Setup: Press the M key
§ Select Application Parameters: Press the
O soft key
§ Select the Totalizing application:
Press the
O soft key
Totalizing
Minimum load for autotaring
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Minimum load for autosave
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autotare first weight
o Off
On
Autostart app when power is on
On
o Off
Autosave mode
o Off
On
Printout when saved
Off
o Print one component on request
Print components
Factory settings
Yes
oNo
o = factory setting
§ Save settings and exit Setup:
Press the M key or the
oo soft key
Minimum Load
The minimum load required for auto-
matic taring of the container weight on
the platform (first weight) is configured
in Setup under:
Application:
Totalizing:
MinL.Autotar
The minimum amount that a compo-
nent must weigh before it can be saved
in totalizing memory is configured in
Setup under:
Application:
Totalizing:
MinL.Autosv
You can choose from the following
10 levels for this setting:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform. If the interval of the con-
nected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (=1000 intervals =
1000 digits) on the weighing platform
for autotaring (only with the “Autotare
first weight” option selected).
Printout
In Setup, under:
Application:
Totalizing: Printout
you can configure whether a printout
is generated manually, by pressing p,
or automatically when a weight value is
stored in the totalizing memory.
If you select
Off for this setting,
printouts can be generated only manu-
ally, by pressing p, and the transac-
tion counter value is not increased.
If the
Print one compo-
nent on request option
is selected, the full printout is generated
every time the “Print” key is pressed;
this printout header lines, date, time,
and dotted line preceding the gross,
tare, net and transaction counter
values.
If
Print components is
selected, the header lines, date, time
and dotted line are included only on
the first printout. Subsequent printouts
include only the gross, tare, net and
transaction counter values.
The total data record is printed
when you clear the totalizing memory
(by pressing the c key).
The total data record includes data
from the gross totalizing memory
(*G), the net totalizing memory (*N),
the total transaction count (n) and
a dotted line.