Operating instructions
37
With the Totalizing application, you
can add weight values to the totalizing
memory. In addition to weight values,
the number of separate values added
to memory is also saved (transaction
counter).
Features
– Weigh in up to 999 items
– Current transaction number displayed in
the text line (indicating the items
already added)
– Weighing in up to a defined target,
with the totalization memory content +
current weight displayed in the text
line.
– Save weight values manually or auto-
matically
– Accurate calculation of total of weight
values from two weighing platforms
– Activate info-mode by pressing D
(> 2 sec)
– Incomplete totalizing routines saved in
battery-backed memory. Configured in
Setup under:
Application:
Totalizing: Autostart
– Automatic taring of container weight.
Configured in Setup under:
Application:
Totalizing: Autotare
– Automatic printout when value saved
– Restore factory default settings.
Configured in Setup under:
Application:
Totalizing: Fty set-
tings
Soft Key Functions
M+ Add net weight value from active
platform to totalizing memory.
The Combics has a totalizing memory
for adding individual net and gross
values. You can save weight values in
totalizing memory manually or auto-
matically. Configured in Setup under:
Application:
Totalizing: Autosave
– Save value manually by pressing M+.
The value taken from the active
platform is added to the value already
saved in totalization memory and
the transaction counter value
is increased by one.
When a value is added manually,
the program does not check whether
the platform has been unloaded since
the last time the
M+ soft key was
pressed.
– Value saved automatically when the
weighing platform is stable and the
defined minimum load is exceeded.
If the defined minimum load is not
exceeded, you can save the item
manually by pressing the
M+ soft key.
Regardless of these settings, the current
value cannot be saved automatically
unless the platform is unloaded before
the next sample is placed on it.
The weighing platform is considered to
be unloaded when the load is less than
50% of the minimum load.
The number of items added to memory
is displayed in the text line.
Press the c key to clear the totalizing
memory. A printout is automatically
generated.
With 2 weighing platforms connected,
you can add values from both platforms
to the totalizing memory.
The displayed result is accurately
calculated in the active weight unit.
Example: When you add 1.243 kg
(determined on a weighing platform
with three decimal places) to 1.4 kg
(determined on a platform with
1 decimal place), the display shows
2.643 kg.
Totalizing s