Operating instructions

28
Weighing in Percent L
Preparation
§ Select Setup: Press the M key
§ Select Application Parameters: Press the
O soft key
§ Select the Weighing in Percent applica-
tion:
Press the
O soft key
Weighing in Percent
Minimum load for autotaring
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Minimum load for initialization
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autotare first weight
o Off
On
Autostart app when power is on
On
o Off
Accuracy for saving weights
o Display accuracy
Display accuracy + 1 decimal place
Display accuracy + 2 decimal places
With internal resolution
Number of decimals for percentages
o None
1 decimal place
2 decimal places
3 decimal places
Parameter for saving weight
o Standard stability
Increased stability
Scale for reference weight
o Do not change
WP 1
WP 2
Display of calculated value
o Residue
Loss
Factory settings
Yes
oNo
§ Save settings and exit Setup: Press the
M key or the
oo soft key
Parameter for Saving Weight Values
The reference weight is saved when
the platform has stabilized. “Stability”
is defined as the point at which fluctua-
tion of a measured value lies within a
defined tolerance range. The narrower
the tolerance range, the more stable the
platform is at “stability”.
In Setup, under:
Application: Wgh in
Pct: Save weight
you can define whether the value
is saved when “standard stability” is
reached, or only at “increased stability”
(narrower tolerance range). If you select
increased stability,
the reference weight saved will be more
accurate and the results more repro-
ducible, but the response time of the
weighing platform might be longer.
Accuracy of Reference Weight
Calculation
The resolution applied for calculating
the reference weight is defined in Setup
under:
Application: Wgh in
Pct: Acc.save wts.
The resolution for calculating the refer-
ence weight is increased if “+1 decimal
place”, “+2 decimal places” or “Internal
resolution” is selected. With the “+1
decimal place” setting, the net value
is determined to one additional decimal
place (i.e., display accuracy x 10);
“+2 decimal places” increases display
accuracy x 100, and so on up to the
maximum resolution available.
Display
With the Weighing in Percent applica-
tion, the result can be displayed as a
remainder or loss. Configured in Setup
under:
Application: Wgh in
Pct: Display
Equations:
Residue = Current weight /
100% weight + 100
Loss = (current weight – 100%
weight) /
100% weight + 100
Minimum Load
The minimum load required for
initialization of the weighing platform
is configured in Setup under:
Application:
Wgh in Pct:
MinL.Init
Once the limit is exceeded by the load,
initialization can begin. If the load on
platform is too light, the following will
occur when you try to save the refer-
ence value:
The error code
inf 29 is displayed
A warning signal is emitted (double-
beep)
The weighing platform is not initialized
The preset reference percentage is saved
The minimum load required for auto-
matic taring of the container weight on
the platform (first weight) is configured
in Setup under:
Application: Wgh in
Pct:
MinL.Autotar
You can choose from the following 10
levels for this setting:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform. If the interval of the con-
nected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (=1000 intervals
=1000 digits) on the weighing platform
for initialization.