Operating instructions
23
Preparation
§ Select Setup: Press the M key
§ Select Application Parameters: Press the
O soft key
§ Select the Classification application:
Press the
O soft key
Classification
Minimum load for autotaring
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Minimum load for Class 1 lower limit
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autotare first weight
o Off
On
Autostart app when power is on
On
o Off
<SET> control output
o ‘SET’ control signal
Ready to operate (for
process control systems)
Activation of port lines
Off
Always on
o Activated at stability
Number of classes
o 3 classes
5 classes
Parameter input
o Weights
Percentages
Auto printout of results
o Off
On
Factory settings
Yes
oNo
o = factory setting
§ Save settings and exit Setup:
Press the M key or the
oo soft key
Minimum Load
The minimum load for the first class
is configured in Setup under:
Application:
Classific.: MinL.Cl1
LL
When this limit is exceeded by the load,
initialization can begin.
Once the application is initialized,
a weight value below the minimum
load is designated Class 0; no class
is displayed.
The minimum load required for auto-
matic taring of the container weight on
the platform (first weight), or for auto-
matic printout of results, is configured
in Setup under:
Application:
Classific.:
MinL.Autotar
You can choose from the following 10
levels for this setting:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform. If the interval of the con-
nected platform is 1 g, for example, and
1000 digits are required, you must
place at least 1000 g (=1000 intervals
=1000 digits) on the weighing platform
for the first class to activate autotaring
or autoprint.
Display:
The result of a given measurement
is shown as either a weight value
or a class number.
– Weight display
The current weight is shown in the
measured value line and the current
class in the text line. Additionally,
the LEDs are lit as follows:
With 3 classes:
– Class 1: red LED
– Class 2: green LED
– Class 3: yellow LED
With 5 classes:
– Class 1: red LED
– Class 2: red and green LEDs
– Class 3: green LED
– Class 4: green and yellow LEDs
– Class 5: yellow LED
– Display of classes
The current class is shown in the
measured value line, and the current
weight in the text line. Additionally,
the LEDs are lit as described above.