Operating instructions

13
With the Averaging application, you
can use your weighing platform for
calculating weights as the average of a
number of individual weighing opera-
tions. These individual operations are
also known as “subweighing opera-
tions.”
This function is used to determine
weights under unstable ambient condi-
tions or for weighing unstable samples
(such as live animals).
Features
Averaging started manually or auto-
matically. Configured in Setup under:
Application: Animal
wgh: Start
With manual start selected, the averag-
ing routine begins when you press a key
(provided the start conditions are met).
With automatic start selected, averaging
begins when you place the first load on
the platform (provided the start condi-
tions are met).
Enter the number of subweighing
operations using the keypad
Info mode
Toggle the display from “result of last
measurement” to “current weight” by
pressing the
Weigh soft key, and
back by pressing
Result
Automatic start of averaging when
the Combics is turned on and a sample
placed on the platform (provided start
conditions are met). Configured in
Setup under:
Application: Animal
wgh: Autostart
Automatic taring of container weight.
Configured in Setup under:
Application: Animal
wgh: Autotare
Automatic printout of results.
Configured in Setup under:
Application: Animal
wgh: Printout
Restore factory default settings.
Configured in Setup under:
Application: Animal
wgh: Factory set
Soft Key Functions
Start Begin animal weighing.
The calculation is made when
the specified number of
subweighing operations have
been completed.
mDef Save the value entered
as “Number of weighing oper-
ations for averaging”.
Begin averaging with the
number of weighing opera-
tions entered as the basis
Weigh. Toggle display to Weighing
mode
Result Toggle display to result
of last measurement
A number of subweighing operations
are required to form the basis for calcu-
lation of an average weight. You can
enter the desired number of subweigh-
ing operations using the keypad.
The number you enter is saved until
it is overwritten by another number.
It also remains in memory when you
switch to a different application
program, or turn off the Combics.
There are three ways to start the
averaging routine:
- Manual start with preset number of
subweighing operations:
Place the sample on the platform and
press the
Start soft key
Manual start with user-defined number
of subweighing operations:
Place the sample on the platform and
enter the number of weighing opera-
tions using the keypad. Press the
mDef soft key to save the number
entered and begin weighing.
Automatic start with preset number
of subweighing operations:
Measurement begins when you place
the first sample on the platform,
provided the start conditions are met.
Averaging (Animal Weighing) V