Operating instructions
8
The weighing platform is powered by the external industrial-grade ING 2 AC
adapter (see “Accessories”). Make sure that the voltage rating printed on this
unit is identical to your local line voltage. If the voltage specified on the label
or the plug design does not match the rating or standard you use, please
contact your local Sartorius office or dealer. To use a main feeder cable from
the ceiling or to mount a CEE plug, you will have to make arrangements
inside your facilities for installation of such cable equipment.
Use only original Sartorius AC adapters, identified by the Sartorius label. Use
of AC adapters from other manufacturers, even if these units have an
approval identification marking from a national testing laboratory, requires the
consent of a trained service technician.
For detailed information on further options for powering the equipment (e.g.,
using local low voltage), contact Sartorius headquarters in Germany or
one of the main offices based in your country.
The equipment must be connected to power in conformance with the
installation requirements of your country.
Safety Precautions
The Sartorius power supply rated to Industrial Class 2 (with safety extra-low
voltage emission (SELV)) can be plugged into any electrical outlet without
additional safety precautions. The ground or earth terminal is connected to
the platform housing, which can be additionally grounded, if required.
Install the power cable with adequate protection to prevent damage to the
cable.
Fixed Power Connection
If the power cable is permanently installed or secured so that it cannot be
unplugged, an emergency switch must be installed for disconnecting power to
the equipment. This emergency switch must be located near the weighing
platform and must be easily accessible. In addition, this switch must be
labeled as an emergency switch. All operators must be instructed in the use
of this switch.
Whenever possible, avoid wiring your weighing platform into networks that
carry a heavy load (such as a connection to a compressor or similar
equipment).