General Specifications
(applicable for use in standard applications, not in legal metrology)
Model IS 16 EDE-H IS 34 EDE-H IS 64 EDE-S IS64EDE-H
Maximum capacity kg 16 34 64 64
Readability g 0.1 0.1 1 0.1
Linearity g ±0.2 ±0.2 ±0.5 ±0.3
Repeatability g
±0.05 ±0.1 ±0.3 ±0.1
(standard deviation)
Response time (average) s <1.5 <1.5 <1 <1.5
Fastest response time (IQ) s <1 <1 <0.8 <0.8
Adaptation to ambient
conditions and application
requirements by selection of one of 4 optimized filter levels
Data output speed 0.1, 0.1, 0.2, 0.4 (normal data output speed)
(depends on filter level selected)
Dust and water protection rating IP65 IP65 IP65 IP44*
Net weight, approx. kg 15 15 15 15
Requirements for power
connection (optional) ING2 (TNG 220/115V)
Frequency Hz 48 - 63 48 - 63 48 - 63 48 - 63
Power consumption:
with isi 10,20,30 VA max. 34 max. 34 max. 34 max. 34
connected to ING2) average 10 average 10 average 10 average 10
Power consumption:
weighing platform only, VA average 2.5 average 2.5 average 2.5 average 2.5
connected to ING2
Selectable weight units g, kg, ct, oz, ozt, tlh, tlt, GN, dwt, mg, /lb, tlc, mom, K, tol, bat, MS
Allowable ambient °C +0 to +40 +0 to +40 +0 to +40 +0 to +40
temperature range (273 K - 313 K) (273 K - 313 K) (273 K - 313 K) (273 K - 313 K)
Allowable operating °C +10 to +30 +10 to +30 +10 to +30 +10 to +30
temperature range (283 K - 313 K) (283 K - 313 K) (273 K - 313 K) (273 K - 313 K)
Automatic zero tracking,
can be turned off by
menu code °C Standard feature Standard feature Standard featureStandard feature
Built-in interface, RS-485 (standard, xBPI protocol): 8 bits; odd parity; transmission speed: 9600–38,400 baud; half-duplex.
Convertible to RS-232 (SBI protocol): 7 bit; parity: even, mark, odd, space; transmission speed:
150–19,200 baud; 1 or 2 stop bits; software/hardware handshake.
Factory setting: 1200 baud; odd parity; 1 stop bit; hardware handshake with 2 characters after CTS
*) = IP 65 (special version available on request)