Operating instructions
13
IS 64 EDE-H0CE (Type of weighing instrument: iso-TEST + BF BF or SARTOCOWAT + BF BF)
verifiable for use in legal metrology.
Specifications group Spec. gp.1* Spec. gp.2 Spec. gp.3 Spec. gp.4
(code): (10 3 1) (10 3 2) (10 3 3)
4
) (10 3 4)
4
)
Maximum capacity kg 64 51 31 16
Readability g 1 1 1 1
Weighing range Standard Configurable Configurable Configurable
Zero tracking±2% kg ±1.3 ±1 ±0.6 ±0.3
Initial zero-setting range kg ±1.3 ±1.3 ±1.3 ±1.3
Preload
1
)kg13264661
isoTEST function
2
) yes yes yes yes
Metrological specifications
Accuracy class kg KKStandard Standard
Minimum capacity
acc. to CD
3
) g 50 50 – –
Minimum capacity
acc. to AWC
4
) g 10000 10000 – –
Verification scale interval g 10 10 – –
Temperature range °C 0 to +40 0 to +40 0 to +40 0 to +40
1
) No reduction in the maximum capacity up to this preload; higher preloads reduce the maximum capacity in 1-kg steps. You cannot set a
preload > approx.14 kg.
2
) The isoTEST function cannot be selected for preloads > approx. 36 kg
3
) CD = Council Directive 2009/23/EC on non-automatic weighing instruments used in the EU and the Signatories of the Agreement on the
European Economic Area
4
) AWC = average weight control based on the requirements of the German Regulations for Prepackaged Products
* = factory setting
IS 64 EDE-H0CEH0CE (Type of weighing instrument: iso-TEST + BF BF or SARTOCOWAT + BF BF)
verifiable for use in legal metrology.
Specifications group Spec. gp.5 Spec. gp.6 Spec. gp.7 Spec. gp.8 Spec. gp.9
(code): (10 3 5) (10 3 6) (10 3 7) (10 3 8) (1039)
Maximum capacity kg 64 64 64 31 16
Readability g 0 to 15 kg: 5 10 20 10 5
15 to 30 kg:10
30 to 64 kg:20
Weighing range Configurable Configurable Configurable Configurable Configurable
Zero tracking±2% kg ±1.28 ±1.28 ±1.28 ±0.6 ±0.3
Initial zero-setting range kg ±1.3 ±1.3 ±1.3 ±1.3 ±1.3
Preload
1
) kg1313134661
isoTEST function
2
) yes yes yes yes yes
Metrological specifications
Accuracy class kg lllll
Minimum capacity
acc. to CD
3
) g 100 200 400 200 100
Minimum capacity
acc. to AWC
4
) g 2500 10000 15000 10000 2500
Verification scale interval g 5/10/20 10 20 10 5
Temperature range °C 0 to +40 0 to +40 0 to +40 0 to +40 0 to +40
1
) No reduction in the maximum capacity up to this preload; higher preloads reduce the maximum capacity in 1-kg steps. You cannot set a
preload > approx.14 kg.
2
) The isoTEST function cannot be selected for preloads > approx. 36 kg
3
) CD = Council Directive 2009/23/EC on non-automatic weighing instruments used in the EU and the Signatories of the Agreement on the
European Economic Area
4
) AWC = average weight control based on the requirements of the German Regulations for Prepackaged Products
* = factory setting