Operating instructions

43
Example:
External calibration and manual adjustment with default weights (factory settings used for weighing parameters)
( Zero the scale
)(press and hold) Start calibration (e.g., when adjustment prompt flashes: WP)
“Cal” is shown for two seconds
You are prompted to place required weight on the platform (e.g., 10,000)
Position the calibration weight on the weighing platform