Operating Instructions Sartorius Combics 3 Models CW3P | CW3S | CW3FS | CH3E | CH3G Complete Combics Scales 98648-011-25
Intended Use Combics 3 is a rugged, easy-to-use complete scale for the complex quality control tasks you perform every day: – – – – in the food industry in the pharmaceutical industry in the chemical industry in the electronics and metal-working industries The Combics 3 scales meet the highest requirements placed on the accuracy and reliability of weighing results, with: – Rugged construction and long service life (stainless steel housing) – Easy operation, thanks to the following features: – large keys
Contents Intended Use 2 Error Codes 64 Contents 3 Warning and Safety Information 4 Care and Maintenance Service Cleaning Safety Inspection 65 65 65 65 Getting Started Equipment Supplied Installation Instructions Transporting the Weighing Platform Leveling the Weighing Platform General View of the Equipment Connecting the Combics to AC Power Connecting the External Rechargeable Battery Pack Connecting a Bar Code Scanner Installing the Verification Adapter for Use in Legal Metrology Operating Toler
Warning and Safety Information The Combics complies with the European Council Directives as well as international regulations and standards for electrical equipment, electromagnetic compatibility, and the stipulated safety requirements. § To prevent damage to the equipment, read these operating instructions carefully before using your Combics scale. ! Do not use this equipment in hazardous areas/locations.
Getting Started The complete scale is available in various versions. If you have ordered special options, the scale is equipped with the required features at the factory. Storage and Shipping Conditions – Allowable storage temperature: –10 ...+40°C (+14 to +104°F) – Once the equipment has been removed from the packaging, it may lose accuracy if subjected to strong vibration. Excessively strong vibration may compromise the safety of the equipment.
Unpacking the Platform Important note: These instructions apply to platforms that are 60 x 80 cm and larger. § Remove the weighing platform and weighing pan from the packaging. § When transporting or lifting the device, hold the unit on the longer sides and observe the safety precautions (wear safety shoes and gloves if necessary). § Remove any plastic wrapping, packaging strips and styrofoam.
Leveling the Weighing Platform Purpose: – To compensate for uneven areas at the place of installation – To ensure that the equipment is placed in a perfectly horizontal position for consistently reproducible weighing results Always level the weighing platform again any time after it has been moved to a different location. § Level the weighing platform using the four leveling feet. Turn the feet until the air bubble is centered in the level indicator.
General View of the Equipment Combics 3 1 2 3 4 1 17 16 15 14 13 12 11 10 2 5 6 7 8 9 3 2 4 5 6 7 8 9 Rear view: CW3P models 10 11 12 13 14 15 16 17 Rear View 18 19 20 22 21 18 21 19 22 23 20 Rear view: CW3S | CW3FS | CH3E | CH3G 18 19 23 22 21 20 8 Load plate Leveling feet Level indicator LEDs for checkweighing and classification Toggle to alphabetical input Alphanumeric keypad CF key (clear function) Settings: Access Setup program Toggle to the application program | application-specific
Connecting the Combics to AC Power § Check the voltage rating and the plug design. $ The scale is powered through the pre-installed power cord. The power supply is built in to the scale, which can be operated with a supply voltage of 100V to 240V. Make sure that the voltage rating printed on the manufacturer’s ID label is identical to that of your local line voltage.
Connecting a Bar Code Scanner (Accessory YBR02CISL) ! Disconnect the equipment from AC power (unplug the AC adapter) § Installation For model CW3P: – Connect a 25-pin D-Sub male connector to the COM 2 (UniCOM) port – To connect both a bar code scanner and an external rechargeable battery, please use the YTC01 T-connector.
For CH3E models: Platform dimensions: 300 x 400 Loading capacity (center) in kg: 130 For CH3G models: Platform dimensions: 560 x 450 Loading capacity (center) in kg: 130 Platform dimensions: 800 x 600 Loading capacity (center) in kg: 600 Shock Resistance The weighing platform features a rugged construction, but you should not allow objects to fall onto the weighing pan. Also avoid bumps to the side of the unit and shocks.
Operating Design With Combics 3 you can – collect weight values from two weighing platforms – use application programs to calculate and display results – assign codes to identify the samples weighed Before you begin, you need to configure your Combics complete scale for your requirements. This is achieved by setting parameters in the operating menu (for example, to configure a connected printer). You can then begin operation, with functions active for storing and calculating weighing data.
Function Keys (Soft Keys) The current function of a given soft key is indicated in the last line on the display (footer). Functions are indicated by abbreviated texts or symbols . 1st ID Input Through the Weighing Platform You can store the weight on the weighing platform; for example, as a tare weight (press the ) key) Numeric Input Through the Keypad § Enter numbers one digit at a time: Press 0, 1, 2... 9 as needed § Store input: Press the required key (e.g.
Operating Design Input Through a Bar Code Scanner or External Keyboard You can use a bar code scanner or a keyboard to enter alphanumeric values in the Combics. Generally, you can use any bar code format that is compatible with the scanner connected.
Bar graph: The bar graph shows the percentage of the weighing platform’s capacity that is “used up” by the load on the scale (gross value). When the Checkweighing application is active, the tolerance limits are also shown (calculated value).
Operating Design Data Output – – – – You can choose from four forms of data output: Printer Digital input/output interface COM port LEDs Printer You can connect one or two strip printers or one or two label printers to the Combics. If you use a universal printer or a model YDP02 or YDP03 printer, you can configure interface parameters in the Setup program (baud rate, stop bits, handshake mode, data bits). The printout can be formatted by the user.
Configuring the Combics Purpose You can configure the Combics 3 to meet individual requirements by entering user data and setting menu parameters in the Setup program. Features The operating menu parameters are divided into the following categories (highest menu level): – – – – – Application parameters Fn key function Device parameters Device-specific information (Info) Language When you use the scale in legal metrology, access to parameters is restricted.
Navigating in the Operating Menu (Examples) Example: Adapting the Combics to ambient conditions at the place of installation; menu item: “Very unstable conditions” for weighing platform WP1.
Defining Password Protection for the Operating Menu: Entering, Changing or Deleting a Password You can define a password to protect the Application parameters and Device parameters. To do this, enter a password known only to authorized personnel. Without the password, only a few of the menu items can be accessed (k key, Info, language).
Printing Parameter Settings To generate a printout of the settings on the current menu level: Press the p key > Printout (example) The maximum width of this printout is 20 characters. -------------------12.01.2002 09:46 Model CW3P1-6ED-LCE Ser.no. 12345678 Vers. 1.1010.10.2 BVers.
Operating Menu Overview (Parameters) o = Factory setting W User-defined setting Setup Application parameters: Please refer to the operating instructions for Combics 3 “Basic Application Programs” Fn-key Off o Gross/net toggling Toggle weight units 10 + higher resolution Device Parameters WP-1 Off RS-232 1) SBI standard SBI verifiable o IS-232 ADC-232 Internal Calibration/Adjustment CAL Key Function o Ext. cal./adjust.; default weight Ext. cal./adjust.
Device Parameters WP-1 Internal Stability Delay No delay o Short delay Average delay Long delay Taring 1) Without stability o After stability Auto Zero o On Off Weight Unit 1 2) User-defined / o (factory setting: grams 1) Grams / g o Kilograms / kg Carats / ct 1) Pounds / lb 1) Ounces / oz 1) Troy ounces / ozt 1) Hong Kong taels / tlh 1) Singapore taels / tls 1) Taiwanese taels / tlt 1) Grains / GN 1) Pennyweights / dwt 1) Milligrams / mg 1) Parts per pound / lb 1) Chinese taels / tlc 1) Mommes / mom 1) Aus
Device Parameters WP-1 Internal Weight Unit 2 2) User-defined / o (factory setting: grams 1) o Grams / g Kilograms /kg Carats / ct 1) Pounds / lb 1) Ounces / oz 1) Troy ounces / ozt 1) Hong Kong taels / tlh 1) Singapore taels / tls 1) Taiwanese taels / tlt 1) Grains / GN 1) Pennyweights / dwt 1) Milligrams / mg 1) Parts per pound / lb 1) Chinese taels / tlc 1) Mommes / mom 1) Austrian carats / K 1) Tola / tol 1) Baht / bat 1) Mesghal / MS 1) Tons / t Display Accuracy 2 o All digits Reduced by 1 decimal plac
Device Parameters COM1 Data Communications o SBI Number of Stop Bits o 1 stop bit 2 stop bits Handshake Mode Software handshake o Hardware, 1 character after CTS Number of Data Bits o 7 bits 8 bits Data Output On request, without stability o On request, after stability Automatic, without stability o 1 display update 2 display updates 10 display updates 100 display updates Automatic, at stability o 1 display update 2 display updates 10 display updates 100 display updates Printout, printer 1 Printout, printer
Device Parameters COM1 Data Communications MP8 Application Program MP8: 3-3-1 MP8: 3-3-2 MP8: 3-3-3 MP8: 3-3-4 MP8: 3-3-5 MP8: 3-3-6 MP8: 3-3-7 MP8: 3-3-8 MP8: 3-3-9 Program Code 2 o Code 2.1 Code 2.2 Code 2.3 Code 2.
Device Parameters COM1 Data Communications SMA Handshake Mode Software handshake o Hardware, 1 character after CTS Number of Data Bits o 8 bits Printer 1 1) YDP01IS o Strip Label Label, manual form feed YDP02 Baud Rate o 1200 baud 2400 baud 4800 baud 9600 baud Parity Space o Odd Even Number of Stop Bits o 1 stop bit 2 stop bits Handshake Mode Software handshake o Hardware, 1 character after CTS YDP03 Baud Rate o 1200 baud 2400 baud 4800 baud 9600 baud 19,200 baud Parity Space o Odd Even Number of Stop Bits
Device Parameters COM1 Printer 1 4) Universal Baud Rate 150 baud 300 baud 600 baud 1200 baud 4800 baud o 9600 baud 19,200 baud Parity Space 1) Odd Even o None 2) Number of Stop Bits o 1 stop bit 2 stop bits Handshake Mode o Software handshake Hardware, 1 character after CTS o YDP04IS o Strip Label Label, manual form feed YAM01IS (external data logger) Printer 2 4) as for Printer 1 External Alibi Memory o YAM01IS COM2 Off WP-2 RS-232: as for COM1 o RS-485 o IS-485 ADC-485 Data Communications o SBI: as for CO
Device Parameters UniCOM (Optional Interface) o Off Data Communications: as for COM1, plus: o SBI: as for COM1 xBPI-232: as for COM1 xBPI-485: as for COM2 SMA: as for COM1 Profibus Valid addresses: from 0 to 126 inclusive; factory setting: 126 Ethernet Optional: Ethernet (for details, see the “Combics UNICOM Interfaces” installation instructions included in this manual) Printer 1: 1) as for COM1 Printer 2: 1) as for COM1 Analog Output Port Value Output o Net value Gross value Error Signal o High (20 mA) Low
Device Parameters Printout FlexPrint o Off On Printer 1 Number of Printouts o 1 printout 2 printouts Components/Individual Printout o Headers 1, 2 o ID1, ... ID4 o Date and time o Application ini data o Scale ID (e.g., serial no.) o Application result Printout of Result/Total o Headers 1, 2 o ID1 through ID4 o Date and time o Scale ID (e.g., serial no.
Setup Device Parameters Operating Parameters Display Backlighting o On Auto shutoff acc. to timer Automatic Shutoff Auto off acc. to timer o No automatic shutoff Timer o 1 + 1 minute 2 + 2 minutes 5 + 5 minutes Main Scale o WP-1 WP-2 Display Geographical Data o Off On Factory Settings: Operating Parameters Yes o No Clock Time: Date: Password Password: Info Service Service date: Terminal Model: Serial no.: Version no.: (application software version) Basic ID: WP1 Model: Version no.
Operating the Combics Weighing W Soft Key Functions The basic weighing function is always accessible and can be used alone or in combination with application programs, such as Counting, Checkweighing, Weighing in Percent, etc.
Operating the Combics Device Parameters Acoustic Signal An acoustic signal is emitted when you press a key. If the key in question is allowed at the time it is pressed, the signal is a single beep. If it is not allowed, a double-beep sounds and the key has no function. You can switch off the acoustic signal in the Setup program, under: Device: Operat.: Acoustic signal Keys You can block the keys to prevent input of key commands. This feature is configured in the Setup program, under: Device: Operat.
Example: Weighing: Tare the scale by placing a container on the weighing platform e Turn on the Combics The self-test function runs. When the display shows a weight readout, the Combics 3 is automatically zeroed and ready to operate. When there is no load on the platform, you can zero the scale at any time by pressing (.
Operating the Combics Example: Weighing: Enter the tare weight value using the numeric keys; print results e Turn on the Combics The self-test function runs. When the display shows a weight readout, the Combics 3 is automatically zeroed and ready to operate. When there is no load on the platform, you can zero the scale at any time by pressing (. .25 Enter the tare weight in the current weight unit using the keypad (in this example, 0.
k Toggle to display of net value p Print the results -------------------24.10.2002 10:09 Model CW3P1-12ED-L Ser.no. 12345678 Vers. 1.1010.10.2 BVers. 01-26-01 -------------------EISENSCHMIDT GOETTINGEN Batch no. 123456 Cust. Smith 24.10.2002 10:09 -------------------G# + 2.250 kg T + 0.000 kg PT2 + 0.250 kg N + 2.000 kg --------------------------------------24.10.
Operating the Combics Example: Weighing: Change the tare values, print results, delete tare values e Turn on the Combics The self-test function runs. When the display shows a weight readout, the Combics 3 is automatically zeroed and ready to operate. When there is no load on the platform, you can zero the scale at any time by pressing (.
Read off net weight p Print the results G# + 6.433 kg T + 4.183 kg PT2 + 0.250 kg N + 2.000 kg -------------------- 0 Enter a zero (“0”) using the keypad ) Save the value entered (all tare values are deleted; the display shows the gross value) p Print the results G# + 6.433 kg T + 0.000 kg N + 6.
Operating the Combics Data ID Codes (Identifiers) Factory settings for the ID code names: This function is available in all application programs and lets you assign codes for identification of measured values (such as product name, batch number, etc.) for inclusion on printouts. ID1: ID2: ID3: ID4: ID1 ID2 ID3 ID4 Features – Up to four ID codes can be stored, edited and individually deleted Factory settings for the ID code values: No default values set. – Each ID code has a name and value.
Example: Entering ID code names. Enter “Batch no.” and “Cust.” as names for ID codes 1 and 2.
Operating the Combics 2 2+ ABC 4+ ABC 2 Soft key l Enter the letter “B”. Press the 2 key repeatedly until the desired character is displayed Enter the letter “a”. Press the 2 key repeatedly until the desired character is displayed. Continue input (in this example, until “Batch no.” is entered) Confirm the name for the first ID code Enter ID code 2 (in this example, “Cust.
Example: Entering ID code values. Enter “123456” and “Smith” as names for ID codes 1 and 2.
Calibration and Adjustment Purpose The accuracy of weighing results must be carefully controlled. This is achieved through calibration and adjustment. Calibration technically means to determine the difference between the weighing instrument readout and the actual weight on the platform to determine the accuracy. This does not involve making any changes in the scale.
Example: External calibration and manual adjustment with default weights (factory settings used for weighing parameters) ( Zero the scale ) (press and hold) Start calibration (e.g., when adjustment prompt flashes: WP) “Cal” is shown for two seconds You are prompted to place required weight on the platform (e.g.
The difference between the weight value and the true mass is displayed, with +/- sign. External calibration Nom. + 10000 g Diff. + 1 g Calibration record is printed, if adjustment was not performed and the process was stopped by pressing ( ) Activate adjustment (press the ( key to cancel) The calibration weight is displayed at the conclusion of calibration -------------------24.10.2002 10:15 Model CW3P1-12ED-L Ser.no. 12345678 Vers. 1.1013.11.2 BVers.
Data Output Functions Data is output on the screen of the indicator and over the interface port. There are two standard interface ports (COM1 and COM 2) and, optionally, a multifunctional interface port (UniCOM). For details, see page 47.
Data in Tare Memory, Calculated Value, Identification of the Active Weighing Platform when More Than One Platform is Used This section shows: a B/G NET PT WP1 WP – Identification of calculated values (values not used in legal metrology) – Identification of gross value or net value (data in tare memory) – Identification of manual tare input (using a bar code scanner) – Display of the active weighing platform when 2 platforms are connected.
Interface Port Purpose The indicator is equipped with the following data interfaces: – Standard COM1 and COM2 interfaces – optional: UniCOM universal data interface (see “Accessories”). Both interfaces can be configured in the Setup program (see “Configuring the Combics”) for different input/ output functions. For example, you can connect a printer, Alibi memory, PC, remote checkweighing display, or second weighing platform to a COM port, or configure the port for control command input (e.g.
COM1 Off * WP2 Data communications Printer 12) or Printer 22) Preparation RS-232 SBI * xBPI-232 MP8 binary SMA YDP01IS YDP02 YDP03 YDP02IS SBI standard version SBI verifiable version IS-232 * ADC-232 Options for Connecting Peripherals You can connect the following printers to the COM1, COM2 or UniCOM port: Strip * Label Label, man. form feed Strip * Label Universal (printer) YDP04IS COM2 External Alibi memory Off * WP21) Data communications Printer 12) or Printer 22) Strip * Label Label, man.
Combics 3 enables connection of a second weighing platform, either to the COM1 port or to the UniCOM universal port. The standard COM1 port is operated in the RS-232 mode. The second weighing platform can be operated in any of the following modes: – SBI – xBPI-232 (factory setting) – ADC-232 – – – – – The standard COM2 port and the optional UniCOM universal interface can be operated in either the RS-232 or RS-485 mode.
Device Parameters Printout Printing Headers Line 1: Line 2: ID codes ID1: ID2: ID3: ID4: ISO/GMP printout o Off For 1 application result For several application results Date/ time o Date with time Date only Once at stability2) o Off On FlexPrint o Off On Printer 1 There are several actions that generate the command for outputting data to the printer port: – Pressing the p key. If the Setup program is active, the menu settings are printed (see “Configuring the Combics”).
This block is not included on the printout of results from the Totalizing and Net-total Formulation applications. – Serial number of the load cell – Results: Gross, tare, and net values; blank line and application-specific result (e.g., piece count) followed by a dotted line. In the Setup menu, select the blocks of information that you wish to include (multiple selections possible; factory settings: all blocks included). A blocked that is not selected is omitted from the printout.
Sample Printouts For details on the individual information blocks, see “Configuring Printouts”, above. For details on configuring the header lines, refer to the “Basic Application Programs” manual for the Combics 3. Weighing Application There is no data for the “initialization data” block. If this block is enabled for the printout, a blank line is output.
Classification Application The “Initialization data” block contains the upper limits of Classes 1 through 4. The “Results” block contains gross, net and tare weights, as well as the class that the sample belongs to (1 through 5, where Class 5 means that the upper limit of Class 4 was exceeded). HEADER LINE 1 HEADER LINE 2 14.07.2002 09:43 -------------------Lim1 + 10.000 kg Lim2 + 11.000 kg Lim3 + 12.000 kg Lim4 + 13.000 kg G# T N + + + 9.700 kg 0.000 kg 9.
Printout of components Example with 3 transactions: HEADER LINE 1 HEADER LINE 2 14.07.2002 09:43 -------------------G# + 1.400 kg T + 0.200 kg N + 1.200 kg n 1 G# T N n + + + G# T N n + + + 3.400 kg 0.200 kg 3.200 kg 2 4.400 kg 0.200 kg 4.200 kg 3 ‘Total’ printout (by pressing c); application data and status as above: HEADER LINE 1 HEADER LINE 2 14.01.
Data Output Format (Line Format) You can output the value displayed in the measured value line and the weight unit, with or without a data ID code. Whether the data ID code is included in the output depends on your settings under “Line Format”. Examples: Qnt + + 235 pcs 235 pcs Without data ID code With data ID code Line Format settings: For raw data (16 characters): no “data header” For other apps. 22 characters): with “data header” (factory setting). Error Codes Pos.
Error Codes Automatic Data Output (SBI) Pos. 1 2 3 4 5 6 7 8 9 10 11 1213 14 15 16 17 18 19 20 21 22 You can have results of measurement printed automatically 1). You can configure the autoprint function to print at certain intervals (measured in display updates 2)) and define whether printing is dependent on stability of the weighing instrument 3) How often the display is updated depends on the operating status and model of the equipment.
Data Input Format You can connect a computer to your indicator to send commands controlling scale functions and applications via the interface port. All commands use the same format, starting with the ESC character (ASCII 27) and ending with a carriage return (CR; ASCII 13) and a line feed (LF; ASCII 10). The total length of a command is anywhere from 4 characters (1 command character between the start and end described above) to 7 characters (4 command characters).
Pin Assignment Charts Model CW3P (IP66/67 Protection) COM1 and COM2 female connectors: Model CW3S: Terminals on the PCB COM1: 11-20 1-10 UniCOM: 1-4 Battery: 1-4 COM2: 1-10 25-contact D-Submini female connector (DB25S) with screw lock hardware Front view Male interface connector used (please use connectors with the same specifications): 25-pin D-Submini (DB25) with integrated shielded cable clamp assembly (Amp type 826 985-1C) and fastening screws (Amp type 164868-1) Pin assignments COM1 (RS-232) Pi
Connections in the CW3S Diagram (on the left): top view Terminal assignments in the 10-contact COM2 terminal strip: RS-232 RS-422 No. 1: Not connected Not connected No. 2: GND GND No. 3: GND GND No. 4: +5 V switch +5 V switch No. 5: Data terminal ready (DTR) Data output - (TxD-) No. 6: Keyboard clock Keyboard clock No. 7: Keyboard data Keyboard data No. 8: Clear to send (CTS) Data input - RxD-) No. 9: Data input (RxD) Data input + (RxD+) No.
Setting the Interface Operating Mode for COM2 Coding for COM2 in Model CW3P RS-232 Solder bridge J2: open Solder bridge J9: open Solder bridge J10: open RS-422 closed open open Coding for COM2 in Model CW3S, CW3FS, CH3E, CH3G RS-232 RS-422 Solder bridge J2: open closed Terminals 5 and 8 on the 10-terminal COM2 strip: open open Terminals 9 and 10 on the 10-terminal COM2 strip: open open 120-ohm terminating resistors for RS-422 and RS-485 RS-422 Solder bridge J4: closed Bias resistors (for RS-485 only) Sol
Connecting Cables to Interfaces Cables should be connected by a certified technician who has received specialized training from Sartorius. ! Make sure to disconnect the equipment from power before connecting cables. ! Installation work that affects the IP67 protection rating must be performed with extreme care. ! Any installation work that does not conform to the instructions in this manual results in forfeiture of all claims under the manufacturer’s warranty.
§ – – – Connect the cable inside the housing as follows: Route the cable through the cable gland. Close and tighten the cable gland in accordance with the applicable regulations. Strip the casing from a section of the cable end (see illustration). The shielding (1) must have contact with the clamps (2). – Expose approx. 15 cm (6 inches) of the individually isolated wires (3) for installation. – Route the cable through the cable gland.
Cabling Diagram (Adapter Cable for PC) Indicator PC V24 — 3 25 pin 2 PC or 9 pin 3 + 4 + — 5 2 20 14 20 4 3 2 5 8 6 6 7 5 7 Cable type: AWG 24 specification 63
Error Codes Error codes are shown on the main display. Err codes are shown continuously; Inf messages are displayed for 2 seconds, after which the program returns automatically to the weighing mode.
Care and Maintenance Service Regular servicing by a Sartorius technician will extend the service life of your Combics scale and ensure its continued weighing accuracy. Sartorius can offer you service contracts, with your choice of regular maintenance intervals ranging from 1 month to 2 years. The optimum length of the service interval depends on the operating conditions at the place of installation and on your requirements. Cleaning !Unplug the equipment from the AC adapter or power supply before cleaning.
Recycling In Germany and many other countries (see www.sartorius.com, Service Download area for details), Sartorius AG or the organization contracted by us takes care of the return and legally compliant disposal of its electrical and electronic equipment on its own. In countries that are not members of the European Economic Area (EEA) or where no Sartorius subsidiaries or dealerships are located, please contact your local authorities or a commercial disposal operator.
Overview Common Specifications Maximum readability Accuracy class Verification scale intervals Digital protective interface Data interface Additional data interface Display Housing: Material Dust and water protection acc. to EN60529 Operating temperature range Power supply DC supply AC supply Battery operation Emissions Immunity to interference Electrical safety 31,250 scale intervals (not in legal metrology) l (for “..-.CE” models) <3000e, (single-range scale) or 2 + 3000e (multiple range scale acc.
Key to Model Designations CH3E Reference device CH3G Information on metrological label Key to model designations for Combics high resolution complete scales Model ID: Load range Readability e=(g), d=(g) Verification scale interval Dimensions Max (kg) Min (g) CH3E -16ED/-H -16ED/-HCE -34ED/-H -34ED/-HCE -64ED/-S -64ED/-SCE -64ED/-H 16 16 34 34 64 64 64 0.1 0.1 0.1 0.1 1 1 0.
Key to Model Designations Here is an example of how to put together order numbers. CW3P1-60 FE-LCE means the following: Complete Combics scale with indicator 3 With one load cell A maximum, single-range weighing capacity of 60 kg A platform size of 500 + 400 mm A resolution of 3,000 e for accuracy class l (CW3P...); (...1...); With selectable application programs.
CW3S4-1500RR-L, our example of a complete, stainless steel scale, means the following: Complete Combics stainless steel scale with indicator 3 With four load cells A maximum, single-range weighing capacity of 1,500kg A platform size of 1,500+1,500 mm A resolution of 15,000 digits (CW3S…); (…4…), (…1500…); (…RR…); and (…L) Complete Combics stainless steel scale With selectable application programs.
CW3S4-1500RR-L, our example of a complete food scale looks like this: Complete Combics Food Scale, stainless steel with indicator 3 (CW3S…), with one load cell, a maximum, single-range (…1…), weighing capacity of 150 kg, (…150…), a platform size of 500 + 400 mm (…FE…) and a resolution of 15,000 digits (…L) Complete Combics Food Scales Applications: weighing, average weight control, counting, checkweighing, classification into 3 or 5 classes, totalizing, net-total formulation, filling, weighing in percent,
Overview Dimensions (Scale Drawings) Indicators: All dimensions are given in millimeters Weighing Platforms: Standard and Stainless Steel Versions Width Height, standard version a b c (mm) (mm) (mm) 320 240 72 400 300 94 500 400 96 650 500 145 800 600 145 800 800 90 1,000 800 90 1,000 1,000 90 1,250 1,000 90 1,500 1,250 90 1,500 1,500 90 2000 1500 100 Length 72 Height, stainless steel version c (mm) 82 101 104 90 90 90 90 90 90 90 90 100 Distance between feet, standard version d e (mm) (mm) 264 184
Dimensions for CH Platforms (Scale Drawings) CH3G CH3E 73
Accessories Indicator accessories: Product Order No.
Product Electrical Accessories External red/green/red display for CW3P models Second display* for CW3P models Remote display*, 7-segment, e. g. 45 mm character size Bar code scanner, 120 mm scanning width, with cable for connection to Combics CW3P models Bar code scanner for Combics CW3S models, adapter cable YCC02-R12F6 required Foot switch, incl. D-Sub 25-pin T-connector Hand switch, incl.
Product Power Supply 24V industrial power supply module External rechargeable battery pack, up to 40 h operation, incl.
Platform Accessories: 800 + 600 800 x 800 1,000 x 800 1,000+1,000 1,250+1,000 1,250 x 1,250 1,500+1,250 1,500+1,500 2,000+1,500 Drive-on ramp, painted Order No. YAR01CWP YAR06CWP YAR06CWP YAR02CWP YAR02CWP YAR03CWP YAR03CWP YAR04CWP YAR05CWP Drive-on ramp, painted (tread plate) Order No. YAR01CWPT YAR06CWPT YAR06CWPT YAR02CWPT YAR02CWPT YAR03CWPT YAR03CWPT YAR04CWPT YAR05CWPT Drive-on ramp, AISI 304 stainless steel Order No.
Product Order No.
Declaration of Conformity C Marking on Sartorius Equipment In 1985, the Council of the European Community approved a resolution concerning a new approach to the technical harmonization and standardization of national regulations. The organization for monitoring compliance with the directives and standards concerning the C marking is governed in the individual EU Member States through the implementation of the EC Directives adopted by the respective national laws.
Index Page Page Accessories Acoustic signal Adjustment/Calibration Bar code scanner, connecting Bar Graph Basic weighing functions Battery, external rechargeable, connecting C marking Cables, installation on interfaces Cabling diagram Calibration/adjustment Cleaning Configuring the Combics Connection to power Data ID codes (ID1 - ID4) Data input format Data output format Data output functions Date and time, setting Declaration of Conformity Device parameters Device-specific information Dimensions Display
General Password M 2 + soft key q, soft key O Activate the Setup program Select Device parameters (or Application parameters) and confirm The password prompt is displayed Enter numbers Soft key l Enter the General Password (see below) Confirm the password The parameter menu is displayed Soft key q (repeatedly, if necessary) Soft key l Read the old password, or enter a new password (max. 8 characters) To delete the password: Press .
Sartorius AG Weender Landstrasse 94–108 37075 Goettingen, Germany Phone +49.551.308.0 Fax +49.551.308.3289 www.sartorius-mechatronics.com Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior written permission of Sartorius AG. The status of the information, specifications and illustrations in this manual is indicated by the date given below.