User manual
7
3.3.4. Reset button
To reset the dispenser, press the reset button on the
battery cover. The display shows E and TIP in blinking
form and pressing the tip ejector button will reset the
pipette to the d-mode.
NOTE: The dispenser should never be reset during
pipetting! Resetting will cause the plunger to move to
the home position and it will empty the contents of the
tip. The tip is also ejected in the end.
3.3.5. Start button
The start button triggers the aspiration and dispensing
operations according to the selected operating mode.
Only a soft touch is required to operate the button.
3. 3.6 . Tip att achment and ejection
Tip attachment
1. Push the tip into the tip connector. Make sure that
the tip is properly attached.
2. Confirm the tip size by double clicking the Start-
button according the instruction on the page 15.
Releasing the tip:
1. Press the Tip Ejection button. The tip is released
immediately.
NOTE: After the dispensing cycle has been completed
a small residual amount of fluid may have been left in
the tip. During the tip ejection it is possible that a part
of the remaining fluid comes out of the tip. Therefore
it is recommended that the tip ejection is performed
above a waste container.
4. Charging of the
Dispenser
The eLINE Dispenser Lite should be charged using the
original Sartorius Charging Stand, Charging Carousel
or AC-adaptor.
1
2
1