User manual
15
NOTE: When aspirating the liquid using the 25 or 50 ml
tip, let the tip stay in the liquid a few seconds after the
plunger has stopped moving. This way you will ensure
that all liquid has been aspirated into the tip.
NOTE: It is possible to work with aqueous liquids
and liquids with higher viscosity. However, it is
recommended always to choose the tip with the
smallest applicable volume in conjunction with high
viscosity liquids. Please note, the specifications are
given in accordance with ISO8655-5.
6. Maintenance
6.1. Cleaning the dispenser
To ensure trouble-free operation it is recommended
to apply regular maintenance and cleaning to eLINE
Dispenser Lite.
To clean and decontaminate the outer surface use
ethanol (70%), isopropanol (60%) or mild detergent and
a soft lint-free cloth. Gently clean the surface of the
pipette with the moistened cloth and wipe dry.
NOTE: Remove the battery pack from the dispenser
prior the cleaning! It is recommended always to use
gloves when cleaning the dispenser. Do not let the
liquid enter the pipette handle!
6.2. Battery replacement
The design of the Dispenser Lite enables fast and easy
battery replacement. The battery replacement pack
includes 4 NiMH batteries fixed to the battery cover. If
the batteries do not hold a sufficient charge for proper
operation follow these steps for replacement of the
battery pack.
1. Hold the dispenser in a horizontal position with the
display facing up.
2. Press the two charging contacts simultaneously to
release the battery pack.
3. Place the new battery pack into the compartment
and close the battery cover by pressing it lightly.