Operating instructions

9
Leveling the Weighing Platform
Purpose:
To compensate for uneven areas at the place of installation
Make sure that the equipment is placed in a perfectly horizontal position for
consistently reproducible weighing results
Always re-level the weighing platform after changing the place of installation.
§ Level the weighing platform using the four adjustable feet. Keep turning the feet
until the air bubble is centered in the level indicator circle.
§ Make sure that all four feet are touching the ground.
> The load must be equally distributed over all leveling feet!
§ Loosen the lock nuts on the leveling feet with an open-end wrench.
> Setting the adjustable feet:
Raise the weighing platform by extending the leveling feet (turning to the right).
Lower the weighing platform by retracting the leveling feet (turning to the left).
§ After aligning the weighing platform, tighten the lock nuts.
Small platforms (1 weighing cell): against the platform frame,
Large platforms (4 weighing cells): against the platform foot.
Operating Tolerances
Never exceed the maximum capacity of the weighing platforms.
The maximum loading capacities of the weighing platforms are listed in the table
below and depend on the position of the weight loaded on the platform (center, sides,
one-sided corner load):
Platform dimensions Center Side Corner
320 x 240 50 35 20
400 x 300 130 85 45
500 x 400 300 200 100
500 x 400 (P*) 600 400 200
650 x 500 (S**) 450 300 150
800 x 600 (P*) 1200 800 400
800 x 600 (S**) 900 600 300
1000 x 1000 4500 3000 1500
1250 x 1000 4500 3000 1500
1500 x 1250 4500 3000 1500
1500 x 1500 4500 3000 1500
2000 x 1500 4500 3000 1500
* Steel
** Stainless steel