Operating instructions
26
Basic
lite
Series and Gem
lite
Series
Model BL310, GM212 BL150, BL6100,
GM312 GM152 GM6101
Weighing capacity g 310 210 150 6100
Readability g 0.01 0.01 0.01 0.1
Tare range (subtractive) g 310 210 150 6100
Repeatability <±g 0.01 0.01 0.01 0.1
Linearity <±g 0.02 0.02 0.02 0.2
Operating temperature range +10…+30°C (50°F…86°F)
Allowable ambient operating
temperature +5…+40°C
Sensitivity drift within <±
+10…+30°C /K 1·10
–5
1·10
–5
2·10
–5
0.5·10
–5
Response time (average) s 2 2 1.5 2
Adaptation to ambient
conditions By selection of 1 of 4 optimized filter levels
Display update (depends on
the filter level selected) s 0.2–0.8
External calibration weight g 100 (F2) 100 (F2) 100 (M1) 5000 (F2)
(of at least accuracy class…) lb* 0.2 – 0.2 10
Net weight, approx. kg
/lb 1.1/2.4 1.1/2.4 1.1/2.4 1.4/3.1
Pan size mm 116 Ø 116 Ø 116 Ø 174x143
inches
4.6 Ø 4.6 Ø 4.6 Ø 6.9x5.6
Dimensions (WxDxH) mm 189x251x70 (7.4x9.9x2.8 inches)
AC power source/power
requirements AC adapter, 230 V or 115 V, +15%…– 20%
Frequency 48–60 Hz
Power requirements,
direct current V 10 to 20
Power consumption (average) W 0.75
Approx. hours of operation with:
– 9-volt battery, approx. h 11
– fully charged
rechargeable battery h 2.5
Hours of operation with the
YRB08Z
rechargeable
battery pack
h25
* = only for BL models