Operating instructions

99
Second connection: For connecting an external battery pack and a bar code scanner
1)
(optional UNICOM interface not installed)
Top view
Terminal assignments in the Terminal assignments in the
10-terminal strip 4-terminal strip
No. 1: Not connected* No. 1: GND_LINE_B
No. 2: GND No. 2: LINE_B
No. 3: GND No. 3: LOW_BATT
No. 4: +5V switch No. 4: BATT_ON_OFF
No. 5: Not connected*
No. 6: Keyboard clock
No. 7: Keyboard data
No. 8: Not connected*
No. 9: Not connected*
No. 10: Not connected*
Connecting Cables to Interfaces
Cables should be connected by a certified technician who has received specialized
training from Sartorius.
! Make sure to disconnect the equipment from power before connecting cables.
! Installation work that affects the IP65/67 protection rating must be performed with
extreme care.
! Any installation work that does not conform to the instructions in this manual will result
in forfeiture of all claims under the manufacturer's warranty.
! Always make sure the equipment is disconnected from power before performing any
installation, maintenance or repair work.
! An IP65/67-protected cable gland for connecting a weighing platform is installed on the
indicator at the factory. Please use extreme caution when performing any work on the
equipment that affects this cable gland. The other openings in the housing are sealed
with protective caps.
Remove the four screws and open the front panel.
ยง Connect the cable from the peripheral device to the indicator.
! A cable gland is installed on the indicator at the factory. Please use extreme caution when
performing any work on the equipment that affects this cable gland. Use a torque wrench
and apply 5 Nm torque to this cable gland.
1)
Combics 2 models only
* depends on the UniCOM used