Operating instructions

96
Error Codes
Position 12345678910111213141516171819202122
Stat*****Er r **##** * * CRLF
or Stat*****Er r *###* * * * CRLF
*: Space
#: Error code number (2 or 3 digits)
ID code characters I
1
Meaning
G# Gross value
N Net weight
T Application tare memory 1
T2 Application tare memory 2
Diff Difference between maximum and minimum
Targ. Exact calibration weight (value)
Nom. Exact calibration weight with SBI output
nRef Reference sample quantity
pRef Reference percentage
wRef Reference piece weight
Qnt Result from Counting (piece count) and
Neutral Measurement applications
mDef Target value for Animal weighing
x-Net Result from Animal weighing
Setp Target value for Checkweighing
Diff.W. Absolute difference (e.g., in kg)
in checkweighing
Lim Deviation in % in Checkweighing
Max Upper limit for Checkweighing
Min Lower limit for Checkweighing
Stat Status
Classx Sorting in classes
Limx Class limits
D Percentage (loss)
Prc Percentage (residue)
Wxx% Reference percentage weight
Cmpxxx Component no. xxx
Cont.T Contents of tare memory
in Net-total Formulation
S-Comp Total weight
in Net-total Formulation
PT2 Preset tare
n Transaction counter
*G Sum of gross weights in Totalizing
*N Sum of net weights in Totalizing
Ser.no Serial number of platform or indicator
1)
depends on scale type; e.g., not all units and characters are
available on scales verified for use in legal metrology.
Automatic Data Output (SBI)
You can have results of measurement printed automatically
1
. You
can configure the autoprint function to print at certain intervals
(measured in display updates
2
) and define whether printing is
dependent on stability of the weighing instrument
3
. How often
the display is updated depends on the operating status and model
of the equipment.
Examples:
N + 153.00 g Net weight
Stat Display blank
Stat L Display underload
Stat H Display overload
Setting:
1) 3)
Automatic output without stability (menu item 6.1.4)
Automatic output with stability (menu item 6.1.5).
Factory setting: Manual with stability;
i.e., automatic output function off.
2)
Time-dependent output:
Intervals: 1, 2, 10 or 100 display updates
(menu item 6.3-x; x=1, 2, 4, or 7).
Factory setting: 6.3.1; 1 display update