Operating instructions

Sorting Application:
The block for initialization data
contains the die upper limits of Classes
1 through 4. The results block contains
gross, net and tare weights, as well as
the class that the sample belongs to
(1 through 5, where Class 5 means that
the upper limit of Class 4 was exceeded).
HEADER1
HEADER2
14.01.2002 09:43
--------------------
Lim1 + 10.000 kg
Lim2 + 11.000 kg
Lim3 + 12.000 kg
Lim4 + 13.000 kg
G# + 9.700 kg
T + 0.000 kg
N + 9.700 kg
Class 1
1)
--------------------
1)
Sample are sorted into 5 classes.
Any weight that exceeds the upper
limit defined for Class 4 is designated
as Class 5, if the application is
configured for 5 classes rather than 3.
Animal Weighing Application:
The block for initialization data
contains the number of measurements
used for averaging. The results block
contains the tare weight and the mean
value.
HEADER1
HEADER2
14.01.2002 09:43
--------------------
mDef 8
T + 0.000 kg
x-Net + 4.202 kg
--------------------
Net/Total Formulation Application:
The block for initialization data is blank.
Which values are shown in the results
block depends on the program
operating status at the time of printing.
The following options are available:
Total/results
(Press c)
Individual/components
(Press O to store component or p
for an individual printout)
93
Total data record
HEADER1
HEADER2
14.01.2002 09:43
--------------------
n3
S-Comp+ 3.400 kg
Cont.T+ 0.200 kg
--------------------
Individual / component printout
If you press O, the header is printed
only once. Each component is printed
automatically when you press O to
store it.
If you are using a label printer, make
sure a single label is large enough for
the list of all components. For printer
models YDP01IS and YDP04IS, you can
configure manual form feed in the
operating menu. With the YDP02IS
printer, form feed is automatic after
each print command (fixed setting).
If an automatic printout is generated
when you store a component, the
component weight is equal to the
current net weight. This is why
components rather than net weights
are printed.
Example with menu setting 3.17.3
active, and 3 components weighed
(the total data record is shown above):
HEADER1
HEADER2
14.01.2002 09:43
--------------------
Cmp001+ 1.200 kg
Cmp002+ 2.000 kg
Print 3rd component by pressing p
G# + 4.400 kg
T + 0.200 kg
T2 + 4.200 kg
N + 0.000 kg
Individual printout when storing
a component in tare memory
(by pressing O).
Example: Print 2nd component
HEADER1
HEADER2
14.01.2002 09:43
--------------------
Cmp002+ 1.000 kg
Individual printout of components
by pressing p
Example: Print 2nd component:
HEADER1
HEADER2
14.01.2002 09:43
--------------------
G# + 2.400 kg
T + 0.200 kg
T2 + 2.200 kg
N + 0.000 kg
Totalizing Application:
There is no data for the “initialization
data" block. If this block is activated
for the printout, a blank line is output.
Which values are shown in the results
block depends on the program
operating status at the time of printing.
The following options are available:
Results printout (press c):
Prints data in gross totalizing memory
(“G"), net memory (“N") and the
number of transactions (“n").
Automatic standard/components
printout (when O is pressed)
Manual standard/components printout
(when p is pressed)
Whether an individual printout or
a printout of all components is
generated depends on the operating
menu settings.
When the components printout is
configured, the header is printed only
once, followed by all components.
If you are using a label printer, make
sure a single label is large enough for
the list of all components. For printer
models YDP01IS and YDP04IS, you
can configure manual form feed in
the operating menu. With the YDP02IS
printer, form feed is automatic after
each print command (fixed setting).