Data Output Functions
There are 3 options for data output:
– Output to the indicator
– Output to the COM1 interface
– Output to a UniCOM multifunction interface (optional)
These two interfaces can be configured in the Setup program for different output
functions, such as connection to a printer or a second weighing platform (WP),
or for data communication with a PC.
With the optional UniCOM multifunction interface installed, you can activate one of the
following functions in addition to the COM1 interface:
– RS-232 interface
– RS-485/RS-422 interface
– Analog output (voltage/current interface)
Output to the Indicator (Weights and Calculated Values)
The indicator is divided into 8 sections. Information about the scale or weighing platform,
the application being used and the sample weighed is output in the following sections:
Plus/minus Sign, Busy Symbol, Zero-setting Range
The following is displayed in this section of the indicator:
– The busy symbol, as long as internal operations are being carried out at the press of a key
– The plus/minus sign of the weight or other measured value
– The zero-setting symbol: identifies “zero” as a weight value (after zeroing the scale or the
active weighing platform)
Line for Measured Values/Results
The following is displayed in this section of the indicator:
– The current weight (on verified scales or platforms with e = d, the last digit is bordered for
identification as a legal value)
– Calculated values (only on Combics 2, e.g., for the Weighing in Percent application)
Unit
The following is displayed in this section of the indicator:
– The current weight unit (such as “g”)
– ID for additional units (only on Combics 2, e.g. for the Counting application)
Data in Tare Memory, Calculated Values, Identification of the Active Weighing
Platform When Several Platforms Are Used
The following is displayed in this section of the indicator:
– Identification of calculated values (values not used in legal metrology)
– Identification of the gross value or data in the tare memory (net value)
– Identification of manual tare input (via bar code scanner) for tare information
– Display of the active weighing platform when 2 platforms are connected (only for
the Combics 2 model)
The display flashes when the weighing platform prompts isoCAL calibration and adjustment
– When the timer is active (menu code 8.9), this symbol flashes to indicate that one-half of
the preset time period has elapsed.