Operating instructions
80
Display
Toggle the display between current
weight and total weight by pressing
w. The numeric display shows the
transaction counter.
When you have toggled to the total
weight display, you can add the current
weight on the weighing platform
to the totalizing memory (manually
by pressing O or automatically,
depending on the Application setting).
Press and hold the key w (> 2 sec) to
toggle to the Info mode. The total of all
gross values is displayed. Press w
again to view the total of all net values.
Press the w key again to show the
transaction counter in the main display
(in this example: n 15 ).
Press the w key again to return to
the original status.
Press and hold the key w again to
exit the Info Mode before the above
sequence is completed.
Printout
In the Application settings you can
configure whether a printout is
generated manually, by pressing p
(3.17.1), or automatically when a
weight value is stored in the totalizing
memory (3.17.2 or3.17.3).
When you generate a printout manually
by pressing p, the transaction counter
value is not printed.
For automatic printouts, you can define
whether a printout is generated after
each individual transaction (3.17.2)
or includes all totalized components
(3.17.3).
Each printout starts with the header
lines and the date and time, followed by
a dotted line, before printing the gross,
tare, net and transaction counter values.
The printout of components includes
the header lines and date and time
only on the first printout. Subsequent
printouts include only the gross, tare,
net and transaction counter values (as
well as the serial number, if configured).
The individual and component printouts
are configured under 7.7 and 7.8.
The total data record is printed when
you clear the totalizing memory. The
total data record includes the data from
the gross totalizing memory (*G, the
net totalizing memory (* N, the total
transaction count (n) and a dotted line.
This printout always includes the lines
configured for Application settings
7.9.x and 7.10.x, regardless of whether
individual or component printing is
configured.
Preparation
§ Press the e key to switch on the
Combics
§ While all segments are lit, press the
) key
§ Open the numeric menu: Press the
) key twice
3 Application parameters
3.5 Minimum load for automatic
taring and automatic printing
3.5.1 1 digit
3.5.2 2 digits
3.5.3 5 digits
3.5.4 10 digits
3.5.5 20 digits
3.5.6 50 digits
3.5.7 100 digits
3.5.8 200 digits
3.5.9 500 digits
3.5.10 1000 digits
3.6 Minimum load for automatic storage
3.6.1 1 digit
3.6.2 2 digits
3.6.3 5 digits
3.6.4 10 digits
3.6.5 20 digits
3.6.6 50 digits
3.6.7 100 digits
3.6.8 200 digits
3.6.9 500 digits
3.6.10 1000 digits
3.7 Automatic taring (first weight tared)
3.7.1 Off
3.7.2 On
3.8 Application started automatically with
most recent initialization data when
Combics switched on
3.8.1 On (automatic start)
3.8.2 Off (manual start)
3.16 Weight stored automatically
3.16.1 Off
3.16.2 On
3.17 Automatic individual or component
printout when value stored
3.17.1 Off
3.17.2 Print individual transaction
by pressing O
3.17.3 Print all components of
transaction by pressing O
§ Save settings and exit Setup: Press and
hold the ) key