Operating instructions
Totalizing s
When the Totalizing application is
selected, you can add weight values
to the totalizing memory. The number
of values added to the memory is also
stored (“transaction counter").
Features:
– Maximum number of values added to
memory: 199
– Values stored automatically after
stability is reached and the minimum
load is exceeded
– Accurate calculation of total of weight
values from two weighing platforms
– Incomplete totalization stored in
battery-backed memory
– Toggle display from current weigh
to total weight: shows the value in
totalizing memory plus the current
weight; for filling to a defined total
– Press and hold w to display the value
in totalizing memory (net, gross) in the
Info mode, accurately calculated in the
active weight unit.
– Toggle the display in Info mode
between total weight and transaction
counter by pressing w; press and
hold w to exit Info mode
Factory settings:
– Minimum load for automatic taring and
automatic printing: 10 digits (3.5.4)
– Minimum load for automatic strorage
of value: 1 digit (3.6.1)
– Automatic taring (first weight on
platform tared): Off (3.7.1)
– Application started automatically with
most recent initialization data when
Combics switched on: Off (manual start)
(3.8.2)
– Weights stored automatically: Off
(3.16.1)
– Print individual/component weight
when value stored in totalizing memory:
Print individual transaction by pressing
O (3.17.2)
79
Operating the Combics 2
The Combics has a totalizing memory
for net and gross values. There are two
ways to store weight values in the
totalizing memory:
– Manually, by pressing the O key.
The value is added to the value already
stored, and the transaction counter
value increases by one.
When a value is added manually,
the program does not check whether
the platform has been unloaded since
the last time the O key was pressed.
– Automatically, when the weighing
platform has stabilized and the
minimum load is exceeded (Application
setting 3.16.2).
The weighing platform is considered to
be unloaded when the load is less than
50% of the minimum load (3.6).
The numeric display shows the
transaction counter.
Press the c key to clear the totalizing
memory. A printout is generated
automatically.
With 2 weighing platforms connected,
you can add values from both platforms
to the totalizing memory. The displayed
result is accurately calculated in the
active weight unit.
Example: 1.243 kg (measured on a
platform using 1 scale interval) added
to 1.40 kg (measured on a platform
using 5 scale intervals) is displayed as
2.643 kg.
Minimum Load
The minimum load at initialization is
configured in the Application settings
(3.5: minimum load for automatic
taring and automatic printing; 3.6:
minimum load for automatic storage
of value). You can choose from the
following ten levels for this setting:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits" here refer to the weighing
platform intervals in the connected
weighing platform. If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g
(= 1000 intervals = 1000 digits) on the
weighing platform for initialization.
If the interval of the connected platform
is 5 g, you have to place at least 5000 g
on the platform for initialization to
achieve the number of digits required in
the above example.
Once the limit is exceeded by the load,
the platform is tared (3.5) or the weight
is stored in totalizing memory (3.6).