Operating instructions
Net-Total Formulation R
When the Net-Total Formulation
application is selected, you can weigh in
different components up to a defined
total. You can also print out the total
weight and the individual weights of
the components.
If the components are filled in separate
containers, please use the Totalizing
application (see page 71).
Features:
– Automatic or manual taring of first
load on the platform
– Minimum load for automatic taring
– Maximum 199 components
– Current transaction number shown
on the numeric display
– Press w to toggle the display:
– In the component mode: Show the
weight of the current component
(shown approx. 1 second after it
is stored, after which the platform
is tared)
– Additive mode: Show the previous
weight plus the current component
weight (shown approx. 1 second after
it is stored, after which the platform
is tared)
– Press and hold the w key to show
precise content of component memory
in the info mode
– Automatic or manual printout of
individual component weights or set of
component weights when value stored
– Toggle to second platform while
calculation is performed for first
platform
Factory settings:
– Minimum load for automatic taring and
automatic printing: 10 digits (3.5.4)
– Minimum load for storing value:
1 digit (3.6.1)
– Automatic taring (first weight on
platform tared): Off (3.7.1)
– Print individual or component
weights when stored: Print the
standard configuration every time
by pressing O (3.17.2)
– Restore factory defaults for Net-Total
Formulation: No (9.1.2)
Minimum Load
The minimum load at initialization is
configured in the Application settings
(3.5: minimum load for automatic
taring; 3.6: minimum load for storing
value). You can choose from the
following ten levels for this setting:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits" here refer to the weighing
platform intervals in the connected
weighing platform. If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g
(= 1000 intervals = 1000 digits) on
the weighing platform for the value
to be stored.
Once the limit is exceeded by the load,
initialization can begin. If the load on
platform is too light for initialization,
the following will occur when you try to
initialize the application:
– The error inf 29 is displayed.
– A warning signal (double-beep)
is emitted
– The value is not stored
Net-Total Formulation with
2 Weighing Platforms
This mode is used for weighing large
and small components at the same
time.
This makes it possible to toggle from
the small-component platform to
the large-component platform during
measurement. Once you toggle to
the large-component platform, the
( and ) keys are available until
a component is stored. For example,
you can tare a partially-filled container
taken from the small-component
platform on the large-component
platform.
The content of the component memory
in the small-component platform is
transferred to the large-component
platform, and the weight value is
converted, if necessary, to the other
weight unit. The Component and
Additive display modes are both
available in the large-component
platform.
The readout from the active platform is
stored in the component memory. The
result is shown in the active weight
unit; for example:
1243 g (from a platform with 1 interval)
+ 1.40 kg (from a platform with 5 intervals)
= 2.643 kg (calculated result)
When you press c to stop a
measurement series, the tare memories
for both platforms are cleared, unless
the large-component platform is an SBI
instrument. In this case, the platform
is only tared.
74