Operating instructions
Installing the Weighing Platform
The connecting cable should be installed by a certified technician who has received special
training from Sartorius.
!Installation work that affects the NEMA rating must be performed with extreme care.
!Any installation work that does not conform to the instructions in this manual will result
in forfeiture of all claims under the manufacturer’s warranty
!Always make sure the equipment is disconnected from AC power before performing
any installation, maintenance or repair work!
!The cable gland (NEMA rating) is already installed on the indicator. Please use extreme
caution when performing any work on the equipment that affects this cable gland!
§ To remove the front panel, remove the 4 screws (see illustration).
§ Use the connecting cable from the weighing platform to connect the indicator.
!The cable gland (NEMA rating) is already installed on the indicator. Please use extreme
caution when performing any work on the equipment that affects this cable gland.
Use a torque wrench. Torque applied for this cable gland: 5 Nm.
§ Strip off the insulation at the cable end and attach the cable as follows:
– Route the cable through the cable gland.
– Properly tighten the screw fasteners of the cable gland.
– Remove the casing from a section of the cable end (see illustration). The shield (1)
must have contact with the clamps (2).
– Expose approximately 15 cm (6 inches) of the wires (3) for installation.
– Route the cable through the cable gland.
– Make sure the shield is in contact with the clamps because the cable is grounded
by the shield.
7