Operating instructions
119
Adjusting the scale:
See also chapter entitled “Operating the Combics", under “Calibration, Adjustment."
Configuring the Combics:
Menu item 1-9-1: External adjustment with default weight,
Menu item 1-10-1: Calibration / adjustment in one procedure,
Menu item 1-16-1: External adjustment / linearization accessible.
( Unload and zero the scale.
) hold Start external adjustment.
This display appears for 2 seconds.
Example: The parameters “elevation of the place of installation" and “geographical latitude"
are entered (menu items 1-20-1 and 1-20-2).
“ALTITUD" is displayed for 2 seconds.
The elevation of the place of installation entered in meters above sea level is displayed.
In this example, the display shows the elevation valid for Germany (“Zone D”).
) Confirm the displayed value or press ( to cancel calibration/adjustment.
"LATITUD" appears for 2 seconds on the indicator.
The geographical latitude of the place of installation entered in degrees
(northern or southern latitude) is displayed.
In this example, the display shows the latitude valid for Germany (“Zone D”).
) Confirm the displayed value or press ( to cancel calibration/adjustment.
The prompt to place the calibration/adjustment weight appears (for example, 50 lb).
Please refer to the chapter entitled “Operating the Combics", under “Calibration/Adjustment"
for more details on the calibration/adjustment procedure.
Instead of “ALTITUD" and “LATITUD", “GRAVITY" is displayed for 2 seconds, if the
acceleration of gravity was entered (under menu item 1-20-3) instead of elevation and
geographical latitude (menu items 1-20-1 and 1-20-2).
The value entered for the acceleration of gravity at the place of installation (in m/s
-2
)
is displayed.
In this example, the display shows the acceleration of gravity valid for Germany (“Zone D”).
) Confirm the displayed value for the acceleration of gravity.
The prompt to place the calibration/adjustment weight appears (for example 50 lbg).
For more details on the Calibration/adjustment procedure, please refer to the chapter
entitled “Operating the Combics", under “Calibration, Adjustment."