Operating instructions

117
Enter of the place of adjustment with subsequent external adjustment (Example)
Setup
(See also the chapter entitled “Operating the Combics", under “Calibration, Adjustment")
§ Remove the cap from the back on the left of the Combics housing.
§ Move the switch to the right (“accessible" position).
§ Service mode setting up (see corresponding section at the beginning of this chapter.
) Select weighing platform “WP1".
) Call up the numeric menu.
Menu page “1" is displayed.
), press k, ) repeatedly Call up menu line 1-13 (tare at power on / zeroing).
The picture shows that tare at power on is “on": The menu item 1-13-1 is selected.
k ) Turn tare at power on “off": Call up menu item 1-13-2.
) ( Return to menu page 1-13.
press
k, ) repeatedly Select menu line 1-20 (Place of adjustment)
Menu item 1-20-1 (degree of latitude) is displayed.
If the geographical latitude and elevation of the place of installation are known, enter this
data by selecting menu item 1-20-1 (degree of latitude) and 1-20-2 (elevation).
To enter the acceleration of gravity at the place of installation select menu item 1-20-3.
Any value entered for the acceleration of gravity under 1-20-3 takes priority over the pair
of values “geographical latitude" and “elevation."
) Example: Enter geographical latitude and elevation.
Call up menu item 1-20-1 (degree of latitude).
In the example shown here, the parameter “acceleration of gravity" (menu item 1-20-3)
has been entered. Therefore, the input fields for the menu items 1-20-1 and 1-20-2
are empty.
In the example shown here, the last geographical latitude entered was the setting
for Germany (“Zone D”). The scale can be used anywhere in Germany with this setting,
if the elevation entered under menu item 1-20-2 is 513 m above sea level.
Enter the required latitude as an absolute number (convert the angular minutes to
decimal places). Press ) to confirm the data entered.
S