Operating instructions
114
If, instead of the A/D converter configuration, the configuration “single range”, “multiple
range” or “multi-interval mode" was selected, then the respective weighing range limit
must be additionally entered into the assigned menu lines (11-5 for “multiple range scale"
and 11-6 for “multi-interval scale"). These menu lines thus contain 3 additional menu
items (the limits for the ranges 1, 2 and 3)
(, press krepeatedly Save configuration: Call up menu line 11-10.
)
Menu item 11-10-2 (“configuration not saved") is pre-configured.
Call up menu item 11-10-1 (“configuration saved").
k
)
Save A/D converter configuration.
The settings for the A/D converter configuration are stored. For a short time, the “Busy"
symbol appears on the indicator, then the program returns to indicator display of menu item,
then Combics is restarted.
After the self-test (i.e. when all segments are displayed), the Combics is in the weighing mode.
Activate the Service mode again (see the corresponding section at the beginning of
this chapter ).
) Select weighing platform “WP1".
) Call up the numeric menu.
Menu “1" is displayed.
), press k ) repeatedly Call up menu line 1-19 (Adjustment without weight).
) Activate menu item “1-19-1".
Enter the nominal capacity of the load cell(s), for example 50.00 lb.
On a weighing platform that consists of several load cells, the nominal capacity must
be multiplied appropriately.
Example: The weighing platform consists of 4 load cells with 50 lb each. In this case,
the nominal capacity (= maximum capacity) 4 + 50 kg = 200 lb.
) Confirms the setting.
S