Depending on the weighing range selected, the relevant menu line is displayed, i.e.
for “single range" (menu item 11-3-1): Input to menu page 11-4,
for “Multiple range scale" (menu item 11-3-2): Input to menu page 11-5,
for “Multi-interval scale" (menu item 11-3-3): Input to menu page 11-6.
The other two menu lines are not accessible.
In the present example , 11-3-1 (single range) is active. This is why the program opened
menu line 11-4 for input. Menu lines “11-5" and “11-6" are not accessible.
Enter the scale interval d:
)
Call up menu item 11-4-1.
) Activate the menu item.
The configuration for the scale interval d is displayed. The flashing number marks the
current cursor position. This marked number can be changed.
The number in the input field can be changed to the desired value by repeatedly moving
the cursor and changing the respective number
Navigation for numeric Input
For more details, see Example 1 and the chapter entitled “Operating Design".
Define the scale interval d for the scale:
If the default value displayed (here: “0.005 lb") corresponds with the desired scale interval,
store the value by pressing the ) key (press for longer than 2 sec.)
Otherwise use the k and ) keys (if necessary, also the ( and p) keys to overwrite
the default value and confirm by pressing the ) key.
In the present example, the scale interval d was set to 0.01 lb.
) Confirms the new setting.
k k k ) Call up the menu item 11-4-4 (maximum capacity).
) enter and save the desired maximum capacity.
The maximum capacity is generally less than the nominal load defined for the load cells
(entered under menu item 1-19-1), as the load cell usually carries additional weight (such as
weighing plattform).
In the present example, a maximum capacity of 30.0 lb was entered.
) Confirms the new setting.