Operating instructions

112
k ) Switch off tare at power on: Call up menu item 1-13-2.
) ( ( Save setting for menu item 1-13-2 and return to menu “1."
k k ) Call up menu 9-1 (factory setting / menu reset).
) k Select A/D converter configuration:
Call up “Standard menu". If the “Trade configuration” is active, the scale cannot be adjusted
without weights by entering the load cell data.
) confirm selection.
The selected configuration is loaded.
For a short time, the “Busy" symbol appears on the indicator, then the program returns
to indicator display of menu item 9-1-2 (“Set default: off", factory setting).
( ( k Select menu page 11:
Configuration: range selection (single range, multiple range scale, multi-interval scale),
scale interval d and maximum capacity Max. cap. (if necessary, also enter range limits).
) k ) Activate menu item 11-3-1
In the present example, the menu item 11-3-1 (“single range") is pre-configured.
If necessary, change the range selection by pressing the k key:
11-3-2: Multiple range scale, 11-3-3: Multi-interval scale.
If necessary, press
k ) to select “Multi-interval scale", confirm selection.
or
k k ) to select “Multi-interval scale", confirm selection.
From more detailed explanations on selecting the range, please refer to the section
entitled “Description of the Individual Menu Items" at the beginning of this chapter.
( Confirm selection and return to the beginning of menu line 11-3.
Enter the scale interval, range limits (only required in the multiple range or multi-interval
k mode) and maximum capacity: