Operating instructions
111
The settings for the A/D converter configuration are stored. For a short time, the “Busy"
symbol appears on the indicator, then the Combics is restarted.
After the self-test when all segments are displayed, the Combics is in the weighing mode.
Adjust the scale: see the chapter entitled “Operating the Combics", under “Calibration,
Adjustment" or the Service Manual.
Following A/D converter configuration, change the menu access switch from
“open" to “locked":
§ if necessary, remove the cap from the back on the left of the Combics housing.
§ Move the switch to the left (to the “locked" position).
See also chapter entitled “Operating the Combics", under “Calibration, Adjustment"
or the Service Manual.
e e Turn the scale off and on again to restart it.
After the self-test, the scale is ready to operate.
Example 2: A/D converter configuration with load cells connected
Setup
(see also the chapter entitled “Operating the Combics", under “Calibration, Adjustment")
§ Remove the cap from the back on the left of the Combics housing.
§ Move the switch to the right (“open" position).
§ Service mode setting up (see corresponding section at the beginning of this chapter).
) Select weighing platform “WP1".
) Call up the numeric menu
) and confirm.
press
k repeatedly , ) Call up menu line 1-13 (tare/zero at power on).
In the picture shown, the tare at power on is set: the menu item 1-13-1 is selected.
S