Operating instructions

110
) Show the configured verification scale interval e.
In accuracy classes l and m the verification scale interval e = d (see also the notes
at the beginning of the section). That is why the program shows this as default value.
( Confirm the verification scale interval e.
k ) When the “Trade configuration" is active, call up the menu item 11-4-3 (minimum capacity)
(if “Multiple-range scale” is set: 11-5-3; with “Multi-interval scale”: 11-6-3).
This menu item is not displayed when the “Standard configuration" is selected.
) When the “Trade configuration" is selected, activate the menu item by pressing ), set the
minimum capacity and confirm the entered value by pressing and holding the ) key (> 2sec).
Example for Class l: Minimum capacity = 20 + e = 2.0 lb
) Confirm the new setting.
k ) Open the next menu item. If “Single-range scale” is set, this is 11-4-4, the last menu item
(maximum capacity).
If “Multiple-range scale” or “Multi-interval scale” is set rather than the A/D converter setting
“Single-range scale”, then the range delimiters must be entered under 11-5 (for the former)
or 11-6 (for the latter). Each of these menu lines has 3 items to choose from, to set limits 1,
2 and 3. The last menu item (11-5-7 or 11-5-8) sets the maximum capacity.
) Enter and save the desired maximum capacity.
In the present example, a maximum capacity of 300.0 lb was entered.
) press for longer than 2 sec. Confirm the new setting.
(, press krepeatedly Save the configuration: Call up menu line 11-10.
)
Menu item 11-10-2 is currently active (“Do not save configuration").
k Select menu item “11-10-1" (“Save configuration").
) Save A/D converter configuration.