Operating instructions
108
) Confirm selection.
If the “Trade" configuration was activated, then at this point the menu line 11-1
(Select Class) is displayed.
Otherwise, i.e. in “Standard" configuration, this menu line is skipped and the menu line
11-2 (1st weight unit) is displayed. In this case, the user must use the k key to select
the menu line 11-3 and proceed as of described behind the figure in menu line 11-3.
) If the Combics is to be configured for use in legal metrology (i.e., the “Trade" configuration
was activated), the accuracy class is already configured (class l)
The setting for menu item 11-1-4 for class l is marked by a circle (
o
) on the indicator.
( k k Select weighing range (menu line 11-3).
) Confirm selection.
In the present example, the menu item 11-3-1 (“single range") is pre-set.
If necessary, change the range selection with k key:
11-3-2: Multiple range scale, 11-3-3: Multi-interval scales.
if necessary,
k ) Select “Multiple range scale", confirm selection.
or
k k ) Select “Multi-interval scale", confirm selection.
For more detailed instructions on how to select the weighing range, see the section on
“Description of the Individual Menu Items" at the beginning of this chapter.
( Return to menu line 11-3.
Enter the scale interval, range limits (only required for multiple range scale or
k multi-interval scale) and the maximum load:
Depending on the weighing range selected, the relevant menu line is displayed, i.e.
for “single range" (menu item 11-3-1): Input to menu page 11-4,
for “Multiple range scale" (menu item 11-3-2): Input to menu page 11-5,
for “Multi-interval scale" (menu item 11-3-3): Input to menu page 11-6.
The other two menu lines are not accessible.
In the present example, 11-3-1 (single range) is active. Therefore, the program opened
the menu line 11-4 for input. The menu lines 11-5 and 11-6 are not accessible.
If the “Trade configuration” is active (class l), then the menu item for setting the scale
interval d is skipped in the submenus 11-4, 11-5 and 11-6 and the second menu item
(for setting the verification scale interval e) is opened, since e = d for these scales.
This menu item is described on the next page.
) Open the first menu item.