Operating instructions
107
A/D Converter Configuration (Examples)
Example 1: Configuring the A/D Converter with a Weighing Platform Connected
Setup
(see also the chapter on “Operating the Combics", section entitled “Calibration, Adjustment")
§ Remove the cap from the back on the left of the Combics housing.
§ Move the switch to the right (toward the interface ports) (“accessible" position).
§ Set to Service mode (see the corresponding section at the beginning of this chapter).
) Select weighing platform “WP1".
) Activate the numeric menu.
Menu page “1" is displayed.
k k ) Activate menu line 9-1.
For navigating in the Operating Menu (Examples) see also chapter on
“Operating Design."
) k Select A/D converter configuration:
Standard configuration (menu item 9-1-3)
if necessary,
k or (press key k again)
Configuration for using the scale in legal metrology (menu item 9-1-4).
) Confirm selected configuration.
The selected configuration is loaded.
For a short time, the “Busy" symbol appears on the indicator, then the program returns
to indicator display of menu item 9-1-2 (“Set default: off ", Factory setting).
( ( k Select menu page “11".
S