Operating instructions

103
1-20 Place of adjustment
(geographical latitude and
altitude, alternatively the
acceleration of gravity at the
place of installation):
1-20-1: Latitude in degrees
1-20-2: Height in meters above
sea level
1-20-3: Acceleration of gravity
1-20-4: Save values for 1-20
9-1 Factory setting / menu reset
9-1-3: Load standard configuration
9-1-4: Load verification configuration
11 Configuring the Analog/Digital
Converter:
11-1 Classes:
11-1-4: Class l
11-2 1st weight unit
(Copy from menu page 1-7):
11-2-1: lo-g to.
11-2-2: Grams
11-2-3: Kilograms
...
11-2-21:Tons
11-3 Range selection:
11-3-1: Single range scale
11-3-2: Multi-interval scale
1
)
11-3-3: Multiple range scale
1
)
11-4 Metrological data for
single range scale:
11-4-1: Scale interval d
11-4-2: Verification scale interval e
11-4-3: Minimum capacity
11-4-4: Maximum capacity
11-5 Metrological data for
Multi-interval scale:
1
)
11-5-1: Scale interval d
11-5-2: Verification scale interval e
11-5-3: Minimum capacity
11-5-4: Range 1
11-5-5: Range 2
11-5-6: Range 3
11-5-7: Maximum capacity
11-6 Metrological data for
Multiple range scale:
1
)
11-6-1: Scale interval d
11-6-2: Verification scale interval e
11-6-3: Minimum capacity
11-6-4: Range 1
11-6-5: Range 2
11-6-6: Range 3
11-6-7: Maximum capacity
1
) Not permitted in legal for trade
applications
11-7 Lock units:
11-7-1: lo-g to.
11-7-2: Grams
11-7-3: Kilograms
...
11-7-21:Tons
11-8 Calibration/adjustment unit:
11-8-1: User-definable unit
11-8-2: Grams
11-8-3: Kilograms
...
11-8-21:Tons
11-10 Save A/D converter
configuration data:
11-10-1:Yes
11-10-2:No*
12-1 Store platform serial number
(verified platform connected
at WP2):
12-1-1: Store serial number
12-1-2: Inactive (standard WP)*
* = Factory setting
Service Menu
When the Service mode is activated, the
service technician can access additional
menu items that are not visible when
the Service mode is deactivated.
The following advanced functions are
available:
In the text menu (Setup):
The following menu items are behind
the menu items Date (“DATE") and Code
(“CODE"):
Service date “S-DATE"
Memory number “MEM-NO"
Serial number of the Combics SER-NO"
In the numeric menu:
The WP1 setup menu (“WP-1") and
the interfaces COM1 (“COM1") and
UniCOM (“WUNICOM") is extended by
the following additional setting options
for configuring weighing platforms
WP-1 (“WP-1") and WP2 (“WP-2"):
1-9 Calibration/adjustment
functions:
1-9-5: Internal linearization
(only for WP-2 connected to
COM1 and UniCOM)
1-9-6: External linearization with
default weights
1-9-7: External linearization with
user-defined weights
1-9-8: Set preload
1-9-9: Clear preload
1-18 Enter the calibration and
linearization weights:
1-18-2: Enter Lin. weight 1
1-18-3: Enter Lin. weight 2
1-18-4: Enter Lin. weight 3
1-18-5: Enter Lin. weight 4
1-19 Adjustment without weight
(Enter the ID codes for the
load cell(s):
1-19-1: Nominal capacity
1-19-2: Resolution
1-19-3: Sensitivity in mV/V for cell 1
(or mean averaged over all cells)
1-19-4: Sensitivity in mV/V for cell 2
1-19-5: Sensitivity in mV/V for cell 3
1-19-6: Sensitivity in mV/V for cell 4
In the case of several cells,
either the individual values are
entered in 1-19-3 ... 1-19-6 or
the mean of all cells in 1-19-3.
1-19-7: Store the values for 1-19