Operating instructions

100
§ Connect the cable as follows:
Route the cable through the cable gland.
Tighten down the cable gland in accordance with the applicable regulations.
Strip the casing from a section of the cable end (see illustration). The shield (1)
must have contact with the clamps (2).
Expose approximately 15 cm (6 inches) of the individually isolated wires (3)
for installation.
Route the cable through the cable gland.
Make sure the shield is in contact with the clamps because the cable is grounded
by the shield.
§ Connecting the wires inside the indicator:
Expose approximately 5 cm (2 inches) of the isolated wires for installation.
Strip approximately 1 cm (1/2 inch) of the isolation from the wires and attach ferrules
to the wires.
Connect the wires securely in accordance with the terminal assignments.
§ After you close the housing again, use a pressure gauge to check the integrity of the
IP65/67 protection. For details, contact the Sartorius Service Center.
Cabling Diagram (Adapter Cable for PC)
(Model CISL1 / CISL2: Adapter cable 7357312, Model CIS1 / CIS2: Adapter
cable YCC02-D9F6) Diagram for interfacing a computer to the indicator using
the RS-232-C/V24 standard and cables up to 15 m (50ft) long.
Cabling diagram
Connection assignments for the cable from the indicator to an RS-232 PC interface
25-contact D-Sub male connector 9-contact D-Sub female
(Model CISL1 / CISL2) connector
1
Sgn GND 7 5 GND
Indicator TxD 2 2 RxD
side RxD 3 3 TxT PC side
DTR 20 8 CTS
CTS 5 4 DTR
Open cable end 9-contact D-Sub female
(Model CIS1 / CIS2) connector
Sgn GND10 5 GND
Indicator TxD 7 2 RxD
side RxD 8 3 TxT PC side
DTR 9 8 CTS
CTS 6 4 DTR
1