§ Connect the cable as follows:
– Route the cable through the cable gland.
– Tighten down the cable gland in accordance with the applicable regulations.
– Strip the casing from a section of the cable end (see illustration). The shield (1)
must have contact with the clamps (2).
– Expose approximately 15 cm (6 inches) of the individually isolated wires (3)
for installation.
– Route the cable through the cable gland.
– Make sure the shield is in contact with the clamps because the cable is grounded
by the shield.
§ Connecting the wires inside the indicator:
– Expose approximately 5 cm (2 inches) of the isolated wires for installation.
– Strip approximately 1 cm (1/2 inch) of the isolation from the wires and attach ferrules
to the wires.
– Connect the wires securely in accordance with the terminal assignments.
§ After you close the housing again, use a pressure gauge to check the integrity of the
IP65/67 protection. For details, contact the Sartorius Service Center.
Cabling Diagram (Adapter Cable for PC)
(Model CISL1 / CISL2: Adapter cable 7357312, Model CIS1 / CIS2: Adapter
cable YCC02-D9F6) Diagram for interfacing a computer to the indicator using
the RS-232-C/V24 standard and cables up to 15 m (50ft) long.
Cabling diagram
Connection assignments for the cable from the indicator to an RS-232 PC interface
25-contact D-Sub male connector 9-contact D-Sub female
(Model CISL1 / CISL2) connector
1
Sgn GND 7 5 GND
Indicator TxD 2 2 RxD
side RxD 3 3 TxT PC side
DTR 20 8 CTS
CTS 5 4 DTR
Open cable end 9-contact D-Sub female
(Model CIS1 / CIS2) connector
Sgn GND10 5 GND
Indicator TxD 7 2 RxD
side RxD 8 3 TxT PC side
DTR 9 8 CTS
CTS 6 4 DTR
1