   CALIBRATION 
 Temperature/CO
2
/O
2
 calibration 
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER 
Key. 
2. Select Temp/CO2/O2 Calibration from the Select Tools Display, press the MENU Key to display the 
menu, select OK, and press the ENTER Key. 
3. The Temp/CO2/O2 Calibration Display will appear.   
Example 
Enter the value shown below if the displayed chamber temperature is 37.0
o
C but the actual temperature 
is 36.5
o
C. 
●Temperature span: Enter 365. 