Welder User Manual
49 
H
2
O
2
 DECONTAMINATION 
When the chamber has been contaminated, or when cleaning the chamber prior to starting a culture, 
H
2
O
2
 decontamination can be performed using the H
2
O
2
 Generator.   
z
 The following products must be purchased separately in order to perform H
2
O
2
 decontamination. Before 
performing H
2
O
2
 decontamination, check to make sure that these are correctly installed. 
  • UV Lamp Expansion Kit (MCO-19UVS) (Not required for the MCO-19M(UV).) 
  • H
2
O
2
 Decontamination Kit (MCO-HL) 
  • H
2
O
2
 Generator (MCO-HP) 
z
 The H
2
O
2
 decontamination function decontaminates the chamber and the inner attachments. Do not 
place any objects other than the specified attachments in the chamber during decontamination. 
z
 Before performing H
2
O
2
 decontamination, clean the chamber and the inner attachments with sterilizing 
alcohol. 
z
 The chamber temperature during H
2
O
2
 decontamination reaches 45°C. Before performing H
2
O
2
decontamination, set the upper limit temperature alarm to 50°C or higher. 
z
 For the H
2
O
2
 reagent, use one bottle of the H
2
O
2
 Decontamination Reagent specified by Sanyo 
(MCO-H202). 
z
 After decontamination, dilute the residual H
2
O
2
 Decontamination Reagent in the H
2
O
2
 Generator with a 
large quantity of water and dispose of it, and rinse with distilled water. (Do not wash either the inside or 
outside of the H
2
O
2
 Generator with alcohol.) 
WARNING
Use the reagent specified by Sanyo for H
2
O
2
 decontamination. Using a different H
2
O
2
 solution may 
cause explosion or damage to the Incubator, or insufficient decontamination. 
Do not use chemicals other than the H
2
O
2
 reagent, such as alcohol. Doing so may result in damage 
to the H
2
O
2
 Generator. 
WARNING 
When performing H
2
O
2
 decontamination, make sure that gastight split doors, inner door and outer 
door are securely closed. During H
2
O
2
 decontamination, plug the access hole with the silicon cap that is 
provided. Failure to do so may be harmful to health due to leakage of H
2
O
2
 gas. 
CAUTION 
H
2
O
2
 decontamination can be performed only for the chamber and chamber attachments with standard 
specifications, and not for any other objects. 