Welder User Manual
31 
6. When gas cylinder A is switched to B, remove gas cylinder A and replace it with new one. 
Note: Be careful when handling the empty N
2
 (or O
2
) gas cylinder. Some gas may still be left in the 
cylinder. 
●When gas cylinder B becomes empty, the line is switched to gas cylinder A again. 
The incubator detects that there is no more N
2
 (or O
2
) gas in a cylinder when the O
2 
density in the 
chamber does not increase for a while even after opening the N
2
 (or O
2
) gas valve in the Incubator. The 
supply line may also be switched for other reasons even though there is gas left in the cylinder: blocking 
or restricting of gas tube, reduction of N
2
 (or O
2
) gas pressure, or improper opening of N
2
 (or O
2
) gas 
cylinder. Always check the gas quantity in the cylinder before disconnecting it.   
● Use the following procedure to manual switch the line between N
2
 (or O
2
) gas cylinders A and B. This 
procedure shows switching the line from N
2
 (or O
2
) gas cylinder B to A as an example. 
1. From the Top Display, press the Up Cursor Key for 3 seconds. 
2. The N
2
 (or O
2
) gas supply line will be switched from cylinder B to A 
  (When the line needs to be switched from N
2
 (or O
2
) gas cylinder A to B, press the Down Cursor Key for 
3 seconds.)   
 5 .
1 
T e m p    3  7  .  0
o 
C   C O 2   5 . 0 %   A B  O 2   5 . 0 %  A  B
2 
3 
4 
5 
    S t a t u s  
6 
  O K     
7 
37.0 
 5.0 
 5.0 
 5 .
1 
T e m p    3  7  .  0
o 
C   C O 2   5 . 0 %   A B  O 2  5 . 0 %  A  B
2 
3 
4 
5 
    S t a t u s   
6 
  O K    
7 
37.0 
 5.0 
 5.0 