Specifications
OWNER’S GUIDE CONDITIONERS, EPAD2/CPAD2 | 16403
Cross-reference of EPAD2 / CPAD2 modules
MODULE INPUT TYPE INPUT RANGES ISOLATION COMMENTS
EPAD2-TH8-x
CPAD2-TH8-x
(where x = J, K, T, U)
8 thermocouple connectors
Type J: xPAD2-TH8-P-J
Type K: xPAD2-TH8-P-K
Type T: xPAD2-TH8-P-T
Type U: UNIVERSAL
Type J: -210 to 1200 °C
Type K: -270 to 1372 °C
Type T: -270 to 400 °C
Type U: includes types K, J, T, E,
R, S, B, N, C, and U
350 VDC (channel to channel
and channel to BUS, power
and chassis)
Overvoltage protection: 15
VDC
Types J, K, T have typical
accuracy of ±0.4°C
Type U (universal) has
typical accuracy of ±1.4°C
EPAD2-V8
CPAD2-V8
8 isolated voltage input
channels
Physical input range: ±50 V
Software selectable: ±100 mV,
±500 mV, ±1 V, ±2.5 V, ±5 V,
±10 V
350 VDC (channel to channel
and channel to BUS, power
and chassis)
Overvoltage protection:
350 VDC
DSUB25 connector - please
order PAD-CB8-B or -BNC
breakout boxes to complete
this module
EPAD2-RTD8
CPAD2-RTD8
8 isolated Resistance
Temperature Detector (RTD)
channels
Resistor: 0 to 999.99 Ω
RTD:
PT100(385), PT200 (385),
PT500(385), PT1000 (385),
PT2000(385), PT100 (3961)
350 VDC (channel to channel
and channel to BUS, power
and chassis)
Overvoltage protection: 15
VDC
LEMO connectors
EPAD2-TH8-P
CPAD2-TH8-P
8 isolated voltage inputs
Supported breakout boxes:
PAD-CB8-x-P2 PAD-CB8-
x-M PAD-CB8-RTD
±1.5 V fixed
350 VDC (channel to channel
and channel to BUS, power
and chassis)
Overvoltage protection: 15
VDC
DSUB25 connector - please
order PAD-CB8-J, K, T or
-RTD breakout boxes to
complete this module
EPAD2-LA8
CPAD2-LA8
8 isolated current inputs 0 to 20 mA, ±20 mA, ±30 mA
350 VDC (channel to channel
and channel to BUS, power
and chassis)
Overcurrent protection:
70 mA cont.
LEMO connectors
Installing CPAD2 modules
After connecting a CPAD2 module to an unused CAN BUS
interface on your Dewetron system, run DEWESoft and go
to the HARDWARE SETTINGS screen, accessed under the
SETTINGS button. Then click on the CAN button.
Select the CAN interface that you connected the CPAD2 mod-
ule to on that screen and set it to work with CPAD2 modules,
as shown in the screen shot here.
If your software does not have this selection on the CAN
screen, please obtain the CPAD2 plugin from Dewetron. This is
a software addon that is free of charge with your CPAD mod-
ules. It was included on a DVD/CD disk when you purchased
your CPAD2 module(s). Or it can be downloaded from our FTP
site.
Additional CPAD2 series modules can be daisy-chained to the
first module. The last CPAD2 module must have the termina-
tor plugged into the last open connector to ensure stable bus
communications.
Terminator