Specifications

16-402 | OWNER’S GUIDE
General EPAD2/CPAD2 Module Specications
Environmental:
Temp. range storage: -30 °C to +85 °C (-22 °F to 185 °F)
Temp. range operating: -5 °C to +60 °C (23 °F to 140 °F)
Enhanced temperature range: Upon request and special order
Rel. humidity (MIL202): 0 to 95 % at 60 °C, non-condensing
RFI susceptibility: ±0.5 % span error at 400 MHz, 5 W, 3 m
All specications within this manual are valid at 25 °C.
All modules are produced according to ISO9001 and ISO14001.
EPAD2/CPAD2 Series Common Information
Calibration information
All DEWETRON modules are calibrated at 25°C after a warmup
time of 30 minutes and meet their specications when leaving the
factory. The time interval for recalibration depends on environmen-
tal conditions. Typically, the calibration should be checked once a year.
Calibration certicates are available from DEWETRON as an option. DEWETRON offers several types:
NIST traceable DEWETRON calibration certicate (USA CAL LAB only)
ISO traceable DEWETRON certicate (European CAL LAB only)
Calibration certicate according to ÖKD (equivalent to DKD)
EPAD2 RS-232/485 interface
EPAD2 series modules are controlled via RS-485 interface, and they require this interface for all data
transfers The DEWE-3210 and DEWE-3211 include an internal RS-232/485 converter and interface. This
converter allows communication with all Dewetron signal conditioning modules. To communicate with the
modules, the RS-232 interface must be set to the following parameters:
baud rate: 9600
data bits: 8
parity: no parity
stop bits: 1
handshake: not required
CPAD2 CAN BUS interface
CPAD2 series modules are controlled via the industry standard CAN BUS over CAN 2.0b protocol, and they
require this interface for all data transfers. CAN BUS is an option for both the DEWE-3210 and DEWE-3211
instruments, and must be installed at the factory in order for this interface to be available for CPAD2 connec-
tion.