Specifications

13-342 | OWNER’S GUIDE
Parameter DAQP-CFB
Over voltage protection ±10 V
Output voltage: ±5V
Output current: ±5 mA
Output protection: Continuous short to ground
Power consumption: max. 1.5 W
Supported TEDS chips: DS2406, DS2430, DS2432, DS2433, DS2431 1)
Weight: within 250 (±30) g
1) TEDS only supported by revision 2 of this module (or higher)
Module Pin-outs
- Pins 6 and 8 can be left unconnected
Software controlled functions
Amplier balance
Amplier balance allows elimination of the amplier offset. The input is shorted and all ranges are balanced
by within the module. Any previously stored sensor offset values are cleared.
Sensor balance
Sensor balance is similar to the amplier balance. Because the input is not shorted, the sensor offset is
automatically adjusted to zero.
Short
When measuring the absolute strain it is possible to disconnect the sensor via software and short the input.
Shunt
Two different internal shunts resistors (50 kΩ and 100 kΩ) can be connected for easy function or calibration
check. With this technique the whole measurement chain (sensor, amplier and analog to digital conversa-