Half bridge signal connection
5-wire sensor (sense wired at the sensor) 3-wire sensor (sense wired at the connector)
Sense leads (SUB-D: pin 3 and 6; LEMO: pin 5 and 6) MUST be connected!
Quarter bridge signal connection
4-wire sensor (sense wired at the sensor) 3-wire sensor (sense wired at the connector)
Sense leads (SUB-D: pin 3 and 6; LEMO: pin 5 and 6) MUST be connected!
1) ‘R+’ has to be connected only if shunt calibration is required, otherwise it can be left unconnected.
Potentiometric and µV measurements
The differential amplier of the DAQP-BRIDGE-B module is designed to measure small voltages (with very
low offset drift and high amplication). These are exactly the same requirements than for μV ampliers.
By setting the bridge input type to Voltage you can select input ranges from ±0.5 mV to ±500 mV. The
advantages of using bridge ampliers for μV measurements: only one multifunctional module with high
bandwidth, a lot of input and lter ranges and a programmable offset (Auto Zero).
The correct hook-up is simply connecting your µV signal to the IN+ and IN- pins of the this module, and be
sure to use a shielded cable and connect the drain to the ground pin of the DSUB connector on the DAQ
module.