Specifications

OWNER’S GUIDE  CONDITIONERS, DAQ | 13313
V
DAQP-HV (and -S3) Isolated High Voltage module (300/700 kHz)
Input ranges: 7 ranges (±20 V to ±1400 V)
Bandwidth: 300 kHz (version DAQP-HV-S3: 700 kHz)
Isolation: 1.8 kVRMS line to line 1.4 kVRMS line to ground
Signal connection: Banana sockets (S3 = Screw terminals)
DAQP-HV Specications
Parameter DAQP-HV
Input ranges unipolar and bipolar: 20 V, 50 V, 100 V, 200 V, 400 V, 800 V, 1400 V
DC accuracy:
20 V and 50 V
100 V to 1400 V
±0.05 % of reading ±40mV
±0.05 % of reading ±0.05 % of range
Gain linearity: 0.03%
Gain drift range: Typically 20 ppm/°K (max. 50 ppm/°K)
Offset drift:
20 V to 100 V
200 V to 1400 V
typical 0.5mV/°K max. 4mV/°K
typical 5ppm/°K max. 20 ppm of Range/°K
Long term stability: 100 ppm/sqrt (1000 hrs)
Input resistance: 10 MΩ
-3dB Bandwidth: 300 kHz
Filter selection: Push button or software
Filter (lowpass): 10, 30, 100, 300, 1k, 3k, 10k, 30k, 100k, 300 kHz
1)
Filter type: Bessel or Butterworth 40 dB/decade
Typical SFDR and SNR:
50 V:
200 V:
1400 V:
300 kHz 100 kHz 10 kHz
SFDR SNR SFDR SNR SFDR SNR
98 76 dB 101 81 dB 108 90 dB
98 84 dB 101 89 dB 108 91 dB
98 86 dB 102 91 dB 107 92 dB
Typical CMRR:
>80 dB @ 50 Hz
70 dB @ 400 Hz
60 dB @ 1 kHz
48 dB @ 10 kHz
Isolation voltage
Line to Ground 1.4kVrms
Line to Line 1.8kVrms
Protection:
Surge (1.2/50)
Burst(5kHz)
CAT III 600
CAT IV 300
±4000V
±4000V
Output voltage: ±5 V
Output resistance: <10 Ω
Output current: 5 mA
Output protection: Short to ground for 10 sec.
Power supply: ±9 VDC ± 1%
Power consumption: 0.7 W