16. Error messages and troubleshooting
Error
message
Possible cause Action
The cuff does not inflate
sufficiently.
The cuff tube is not
connected correctly to the
device.
Attach the plug on the cuff tube
correctly to the device.
Tube or cuff leaky. Replace the
tube or cuff.
Put the cuff on correctly and
take the measurement again.
Er2 The cuff is not in the right
position on the arm.
Put the cuff on correctly and
take the measurement again.
Er3 Strong vibration during
measurement.
Keep still and do not speak
during measurement.
Er5 Measurement has failed
because you moved or spoke.
Keep still and do not speak
during measurement.
Er7 Measurement failed. Put the cuff on correctly and
take the measurement again.