User Manual
Table Of Contents
- CONTENTS
- Chapter 1 Product Overview
- Chapter 2 Working Principle
- 2.1 Microwave/Millimeter-Wave Frequency Conversio
- 2.2 Frequency Synthesis Module
- 2.3 IF Signal Processing Module
- 2.4 Data Processing, Display and Control Module
- Chapter 3 Analysis on Common Faults
- 3.1 Lighting failure of power indicator lamp
- 3.2 No response of key
- 3.3 No zero frequency signal
- 3.4 Frequency accuracy out of standard
- 3.5 No signal detected when the pan smaller than
- 3.6 Large deviation between the signal amplitude
- condition
- 3.7 SWR deviation at input port
- 3.8 Phase noise deviation
- 3.9 Abnormal attenuation control
- 3.10 No display for battery under charging and cu
- Chapter 4 Repair of Parts
- 4.1 Removal Process
- Chapter 5 Calibration of the Main Unit
- 5.1 Temperature Compensation
- 5.2 Automatic Test of Indicators
Tel: 886.909 602 109 Email: sales@salukitec.com
www.salukitec.com
14
4. Remove the rear casing assembly:
Firstly, remove
and replace the frequency synthesis board (12):
a) Firstly, remove 6 fixing screws of the frequency synthesis board (12) in the rear cover assembly (19).
b) Loosen 3 semi-rigid cables between the frequency synthesis board (12) and the
microwave frequency conversion module (13).
c) Remove the interconnecting flat cable between the microwave frequency conversion module (13) and the
frequency synthesis board (12).
d) The installation process is reverse to the removal process.
Fig.6 Removal of the frequency synthesis board
The microwave frequency conversion module (13) after removal of the frequency synthesis board (12) is as
shown in the following figure. The removal and installation processes are as follows:
a) Remove 4 fixing screws between the microwave frequency conversion module (13) and the rear cover
assembly (19). Please note that two screws on the right side have two washers.
b) Remove the connection flat cable between the attenuator (15) and the microwave frequency conversion
module (13).
c) Remove the connection semi-rigid cable between the attenuator (15) and the microwave frequency
conversion module (13).
d) After the module is removed, the rest is the rear cover assembly (19).