Chapter 4 Repair of Parts
When the S3302 series spectrum analyzer has any fault and the casing shall be opened to check the fault cause or
the spare parts and complete parts shall be replaced, please refer to the schematic diagram of S3302 structure
decomposition, and remove and install it according to the instruction.
Fig.2 S3302 exploded view
4.1 Removal Process
1. Remove the battery cover assembly (17), and then remove the battery (11).
Fig.3 Removal of battery
2. Gently place the S3302 series spectrum analyzer on the table with back upwards, and remove eight fixing
screws (four on the back and four on the side) of the casing with a Phillips screwdriver.