7 Sweep menu
·[Sweep Time Auto Man]: press [Sweep]
[Sweep Time Auto Man], to
switch the sweep time between automatic and manual modes. The current
state will be underlined. When the sweep time is automatic, the integrated
tester will automatically set the fastest sweep time as per the current
apparatus state and the fastest sweep time will be displayed on screen.
When the sweep time is manual, it is allowed to enter the sweep time by
using number key and select the time unit to complete setting by using
relevant soft key. In case of automatic sweep time, the scan frequency
speed will vary with RBW and VBW. The bigger RBW and VBW are, the
quicker the scan speed is; the less RBW and VBW are, the slower the
scan speed is. When the minimum sweep time limit is met, the sweep time
of S3302 series spectrum analyzer can set up to 800s for non-zero span
and 600s for zero span.
·[Sweep Type Cont Single]: press[Sweep][Sweep Type Cont Single],
to select the sweeping mode. The setting of sweep type can decide the
scanning direction of an integrated tester and the time when it can stop
scanning and enter the hold mode. Under the interference analysis mode,
there are two sweeping modes, continuous sweeping and single sweeping.
·[Sweep Once]: press [Sweep][Sweep Once], to start single again.
·[Points]: press [Sweep][Points], to select the scan points as [201],
[501], [1001], [2001] and [4001].