143
6 Peak menu
· [Maximum Peak]: Press [Peak]→[Max Peak] to set the current
active marker at the maximum peak of the measured trace. The
frequency and amplitude of the marker will be displayed in the
middle upper part of the screen.
· [Next Peak]: Press [Peak]→[Next Peak] to set the active marker
at next peak associated with the current marker location on the
trace.
· [Next Peak Left]: Press [Peak]→[Next Pk Left] to find next peak
on the left side of the current marker location.
· [Next Peak Right]: Press [Peak]→[Next Pk Right] to find next
peak on the right side of the current marker location.
· [Maximum value]: Set one marker at the highest point of the
trace. The frequency and amplitude of the marker will be displayed
in the right upper corner of the screen. The active functions will not
change when this key is pressed.
· [Minimum value]: Press [Peak]→[Min Search] to set one marker
at the lowest point of the trace. The frequency and amplitude of the
marker will be displayed in the right upper corner of the screen.
The active functions will not change when this key is pressed.
· [Marker→Center]: Press [Peak]→[Marker→Center] and set the
center frequency as the marker frequency. By using this function,
the signal can be rapidly moved to the screen center, which is only
effective for the RF spectrum.
· [Special note]: Each peak, maximum value and minimum
value will be searched on the selected trace. Refer to [Select
Trace] menu for details.