Technical information
Installation
Doc. 065186-02 1/12 15
For a Dionex ICS-1000/1500/2000 pump:
a. Highlight relay 2 on the State Devices tab page.
b. Press the F2 key to display the Device Configuration dialog box.
c. Select Flow Zero as the Function for relay 2. This ensures that
the Dionex VP is automatically turned off whenever the
chromatography pump is turned off or the flow rate is set to zero.
NOTE The Flow Zero function is not available unless
Dionex ICS-1000/1500/2000 Moduleware
version 3.0.0 (or later) is installed.
d. Click OK.
8. Click OK to close the Properties dialog box.
9. If it is not already running, start the Chromeleon client. (Select Start
> All Programs (or Programs, depending on the operating system) >
Chromeleon > Chromeleon.)
10. In the Program Wizard, specify when to start and stop the relay
selected in Step 7
.
For example, the Relay and State Devices Options tab page for the
Dionex ICS-3000 DC is shown here.
If the Dionex VP continues to draw a vacuum when the
chromatography pump is not pumping liquid, the consumable or
accessory may be damaged. For example, the vacuum may destroy
the membrane inside the Dionex CRD.