Technical information

Installation
Doc. 065186-02 1/12 9
7. If you plan to operate the Dionex VP in the relay control mode (see
page 1
), check that the RELAY CNTRL button is in the ON position (the
button is pressed in). Then, complete these additional steps:
a. Locate the relay connector (P/N 923617) and twisted wire
assembly (P/N 043598) in the Thermo Scientific Dionex VP
Vacuum Pump Package (P/N 066463).
b. Insert the ends of the wires into the plug. (If necessary, strip the
ends of the wires first.)
c. Connect a red signal wire to the + pin on the Dionex VP. Connect
a black ground wire to the - pin.
d. Connect the other ends of the signal wires to the corresponding
relay output signal and ground pins on the controlling device. For
instructions on how to connect the Dionex VP to a Dionex ICS-
3000 DP/SP, go to page 9
. For instructions on how to connect the
Dionex VP to a Dionex ICS-1000/1500/2000, go to page 10
.
Connecting the Dionex VP to a Dionex ICS-3000 DP/SP
1. Locate the 25-position
connector plug
(P/N 923968) in the
Dionex ICS-3000
DP/SP Ship Kit.
Remove the cover of
the plug (see Figure 5
).
Figure 5. 25-Position Connector Plug (Cover removed)
Position 20
Position 15
Position 14
Position 8