Installation manual
Autosamplers
www.thermo.com
63
Thermo Scientific
FS95 Furnace
Autos
ampler
The attractively styled Thermo Scientific FS95 Furnace Autosampler will accommodate up to
60 samples in the large carousel. Carousel changeover allows up to 500 samples per run.
Six (6) large reagent cups enable up to six (6) matrix modifiers to be employed. Using the cup
reducing ring, smaller sample cups can be used in the reagent cup positions to reduce modifier
consumption. The sample is collected and dispensed via an inert PTFE capillary with rapidly
interchangeable tips. Viscous and normal modes of injection ensure that samples are handled
appropriately, and an enhanced wash program eliminates contamination. Fast furnace operation
provides concurrent operation of the furnace, spectrometer, and autosampler sample uptake
procedures, minimizing overall furnace cycle time.
The syringe is visible and easily accessible for routine maintenance. A large, 1-liter capacity
wash vessel for long, unattended analysis is complemented by an on-board waste vessel,
removing the need for inconvenient waste collection on the laboratory floor. All facilities are
programmed from the SOLAAR software and sampling facilities. Matrix modification (wet or
dry mixing), standards addition, dilution, re-concentration and automatic standards preparations
can be rapidly set up. Automatic re-scaling and recalibration functions are provided. If a sample
falls outside the calibration range the system automatically and intelligently calculates the
most appropriate dilution factor required to bring the sample back within range. Autosampler
loading guides indicate the optimum carousel layout.
Standard Items
• Furnace Autosampler
• Polypropylene Sample Cups (1000)
• Polypropylene Reagent Cups (50)
• Spare Capillary Tips (10)
Necessary Additional Items: None
Carousel Capacity: 60 sample cups, 6 reagent cups
Cup T
ypes: Polypropylene sample and reagent, fluoroplastic sample, reduced volume sample
Cup Volume (nominal): Sample 2 mL, reduced volume 1.5 mL, reagent 25 mL
Matrix Modifiers: Up to 6 per method
Resamples: 1 to 99 per measurement
Sampling System: 100 µL syringe (side port)
Sampling Volume: 1 to 70 µL
Reproducibility of Injected Volume: Better than 1% (by mass) at volumes => than 10 µL
Accuracy of Injected Volume: ±0.5 µL
Inert Gas Pressure
• Nominal – 0.34 bar (5 psi)
• Minimum – 0.20 bar (3 psi)
• Maximum – 0.68 bar (10 psi)
W
ash and Waste Container Volume: 1 Liter each
Dimensions (mm): 296 W x 380 H x 320 D
Weight: Approximately 7.3 kg
Ordering Information
Description Part Number Price
FS95 Furnace Autosampler (M Series Only) 942348020001 CALL