Service manual
14 SiteStand Mobile Docking Station Service Manual
3.4 SiteStand Mobile Docking Station
Performance Tests
3.4.1 Setup
1. Install any SonoSite ultrasound system (without a battery) into the SiteStand
mobile docking station.
2. Apply AC power, 120V 60Hz to SiteStand station “AC in” connector.
3.4.2 AC Power Verification
Use an AC Tester to verify both IEC output connectors are wired correctly.
3.4.3 Docking Video
1. Turn on the system.
2. Connect a Test Monitor to each of the video outputs on the docking stand.
3. Verify the same video pattern exists on the Test Monitor as is present on the
System LCD Display.
3.4.4 RS-232
The RS-232 interface can only be field tested using the SiteLink Image Management
software. If that software is installed, verify the information on the PC according to
the SiteLink Image Management User Guide on the SiteLink CD-ROM.
3.4.5 Printer Interface
1. Connect the black and white video printer to the SiteStand station. Ensure that
the control and AC power cables are attached, but not the video cable.
2. Verify that the printer produces an audible beep and that the user interface
replies with its own tone.
3. Repeat steps 1 and 2 (above) at least one more time to verify that the status line
is connected.
3.4.6 Audio Tone
1. Connect an amplified speaker to the audio output connection on the stand.
2. Set the system to Doppler function and perform a scan on a live model or a
Doppler phantom.
3. Verify that the audio output to the amplified speaker is comparable to that
produced by the system speakers.